Contactless Detection of pH Change in a Liquid Analyte
We describe an experiment in which we employ a radiofrequency sensor to measure pH changes in a liquid solution. The experiment is novel in a few ways. First, the sensor does not have contact with the liquid but rather detects the change from the outside of a PVC pipe. Second, the change is detected using a Linear Discriminant Analysis model using values from an inverse Fourier transform of the frequency data as its features. We believe this to be the first use of Fourier analysis in contactless pH measurement using radio frequencies.
Authors
- Dominic Klyve (ORCID: https://orcid.org/0000-0002-4424-3130)
- Debbie Gustafson
Institutions
- Central Washington University (US)
- Patient Knowhow (United States) (US)
Publication Details
- Journal
- Sensors
- Published
- 2025-04-30
- DOI
- https://doi.org/10.3390/s25092832
- Primary Topic
- Analytical Chemistry and Sensors
- Type
- article
- Field-Weighted Citation Impact
- 0.00