A systematic investigation on the response of EBT-XD gafchromic films to varying dose-per-pulse, average dose-rate and instantaneous dose-rate in electron flash beams

Background The FLASH effect is a radiobiological phenomena characterized by the reduction of the damage to healthy tissues while maintaining iso-effectiveness on the tumour, obtained by delivering the entire radiation dose in less than 100–200 m and with average dose rate greater than 40–100 Gy/s. Despite the enormous interests of the scientific community, a series of issues must be addressed and overcome to reach its clinical implementation. One of these challenges is related to the dosimetry of ultra-high dose-per-pulse (UHDP) beams, which trigger the effect. The most used dosimeters to date in radiobiological experiments with ultra-high dose rate (UHDR) beams are the radiochromic films. However, a systematic and accurate study of their response by varying dose-per-pulse and dose rate over a wide range has never yet been done. Purpose To systematically investigate the response of EBT-XD radiochromic films under UHDR electron beam irradiations by varying different beam parameters independently and over a wide range and using flashDiamond as reference. Materials and Methods Thanks to a special research linac, average dose rate (ADR), dose-per-pulse (DP) and instantaneous dose rate (IDR) dependencies on film response have been individually investigated. The reference value of ADR, DP and IDR has been calculated by measuring the dose using a flashDiamond and knowing the temporal beam structure provided by the beam monitoring system of the Linac. Results Our results normalized to fD showed an overresponse of the radiochromic films, compared to a conventional irradiation, for ADR values higher than 1000 Gy/s and an IDR higher than 1 MGy/s which could represent a total dose overestimation of about 10%. Discussion The gafchromic film EBT-XD have their field of application in UHDP beam dosimetry, considering their excellent spatial resolution; However, particular attention must be paid when using this type of dosimeter for absolute dose measurements at very extreme values of IDR (>1 MGy/s).

Authors

Institutions

Publication Details

Journal
Frontiers in Physics
Published
2025-04-29
DOI
https://doi.org/10.3389/fphy.2025.1474416
Citations
11
Primary Topic
Radiation Effects in Electronics
Type
article
Field-Weighted Citation Impact
8.03
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A systematic investigation on the response of EBT-XD gafchromic films to varying dose-per-pulse, average dose-rate and instantaneous dose-rate in electron flash beams

Jake Harold Pensavalle, Alessandra Gonnelli, Francesco ROMANO, Giovanni Gadducci et al.
11 citations
Frontiers in Physics
Radiation Effects in Electronics
8.03
article

A systematic investigation on the response of EBT-XD gafchromic films to varying dose-per-pulse, average dose-rate and instantaneous dose-rate in electron flash beams

Jake Harold Pensavalle, Alessandra Gonnelli, Francesco ROMANO, Giovanni Gadducci, T. Fuentes, L. Masturzo, D. Del Sarto, Fabio Di Martino, Noemi Giannini, G. Milluzzo, Fabiola Paiar, Maria Celentano, A. L. Cavalieri
article en
11 citations

Abstract

Background The FLASH effect is a radiobiological phenomena characterized by the reduction of the damage to healthy tissues while maintaining iso-effectiveness on the tumour, obtained by delivering the entire radiation dose in less than 100–200 m and with average dose rate greater than 40–100 Gy/s. Despite the enormous interests of the scientific community, a series of issues must be addressed and overcome to reach its clinical implementation. One of these challenges is related to the dosimetry of ultra-high dose-per-pulse (UHDP) beams, which trigger the effect. The most used dosimeters to date in radiobiological experiments with ultra-high dose rate (UHDR) beams are the radiochromic films. However, a systematic and accurate study of their response by varying dose-per-pulse and dose rate over a wide range has never yet been done. Purpose To systematically investigate the response of EBT-XD radiochromic films under UHDR electron beam irradiations by varying different beam parameters independently and over a wide range and using flashDiamond as reference. Materials and Methods Thanks to a special research linac, average dose rate (ADR), dose-per-pulse (DP) and instantaneous dose rate (IDR) dependencies on film response have been individually investigated. The reference value of ADR, DP and IDR has been calculated by measuring the dose using a flashDiamond and knowing the temporal beam structure provided by the beam monitoring system of the Linac. Results Our results normalized to fD showed an overresponse of the radiochromic films, compared to a conventional irradiation, for ADR values higher than 1000 Gy/s and an IDR higher than 1 MGy/s which could represent a total dose overestimation of about 10%. Discussion The gafchromic film EBT-XD have their field of application in UHDP beam dosimetry, considering their excellent spatial resolution; However, particular attention must be paid when using this type of dosimeter for absolute dose measurements at very extreme values of IDR (>1 MGy/s).

Frontiers in PhysicsVol. 13
University of Pisa (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Catania (IT), Presidio Ospedaliero (IT), Azienda Ospedaliera Universitaria Pisana (IT)
Openalex Percentile: Top 2%
Radiation Effects in Electronics
8.03
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.