Evaluasi Dosis Radiasi Hambur Pemeriksaan CT Scan Thorax pada Organ Gonad dan Tiroid

Research has been carried out regarding the evaluation of scattered radiation dose during CT scans of the thorax on the gonads and thyroid using 30 patients with CT scans of the thorax. In this study, the radiation dose received by thorax CT Scan patients was measured using TLD-100 which was placed at 3 measurement points on the patient. The first point was placed on the thyroid using 1 TLD, the second point on the thorax using 2 TLDs, and the third point was placed on the gonads using 1 TLD. The thorax CT Scan examination used a tube voltage of 120 kV. The results showed that the DLP and CTDIvol values ​​affected the value of the effective dose received. The average values ​​of CTDIvol and DLP were 6.268 mGy and 782.196 mGy.cm and the average thorax dose was 2.495 mSv. The average CTDIvol and DLP values ​​obtained in the study did not exceed the threshold set by the BAPETEN Regulation in 2021, namely the CTDIvol and DLP values ​​were 16 mGy and 810 mGy.cm. Radiation that affects the gonads and thyroid organs is scattered radiation. The average radiation dose received on the gonads and thyroid organs was 0.121 mSv and 0.485 mSv. The average dose value received did not exceed the dose received on the thorax and the radiation dose received on the thyroid was greater than the radiation dose received on the gonads. This is because the thyroid organ is closer to the irradiation area than the gonad organ.

Authors

Publication Details

Journal
Jurnal Fisika Unand
Published
2025-04-29
DOI
https://doi.org/10.25077/jfu.14.1.82-88.2025
Primary Topic
Edcuational Technology Systems
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluasi Dosis Radiasi Hambur Pemeriksaan CT Scan Thorax pada Organ Gonad dan Tiroid

Syarifatul Ulya, Rico Adrial, Magfirah Magfirah
Jurnal Fisika Unand
Edcuational Technology Systems
article

Evaluasi Dosis Radiasi Hambur Pemeriksaan CT Scan Thorax pada Organ Gonad dan Tiroid

Syarifatul Ulya, Rico Adrial, Magfirah Magfirah
article en

Abstract

Research has been carried out regarding the evaluation of scattered radiation dose during CT scans of the thorax on the gonads and thyroid using 30 patients with CT scans of the thorax. In this study, the radiation dose received by thorax CT Scan patients was measured using TLD-100 which was placed at 3 measurement points on the patient. The first point was placed on the thyroid using 1 TLD, the second point on the thorax using 2 TLDs, and the third point was placed on the gonads using 1 TLD. The thorax CT Scan examination used a tube voltage of 120 kV. The results showed that the DLP and CTDIvol values ​​affected the value of the effective dose received. The average values ​​of CTDIvol and DLP were 6.268 mGy and 782.196 mGy.cm and the average thorax dose was 2.495 mSv. The average CTDIvol and DLP values ​​obtained in the study did not exceed the threshold set by the BAPETEN Regulation in 2021, namely the CTDIvol and DLP values ​​were 16 mGy and 810 mGy.cm. Radiation that affects the gonads and thyroid organs is scattered radiation. The average radiation dose received on the gonads and thyroid organs was 0.121 mSv and 0.485 mSv. The average dose value received did not exceed the dose received on the thorax and the radiation dose received on the thyroid was greater than the radiation dose received on the gonads. This is because the thyroid organ is closer to the irradiation area than the gonad organ.

Jurnal Fisika UnandVol. 14(1)
Openalex Percentile: Top 96%
Edcuational Technology Systems
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Evaluasi Dosis Radiasi Hambur Pemeriksaan CT Scan Thorax pada Organ Gonad dan Tiroid — Syarifatul Ulya, Rico Adrial, et al. · Jurnal Fisika Unand (2025) | TGRS Research Map | TGRS