Preliminary Evaluation of an Automated Blood Cell Analyzer for Its Use with Blood Samples from Rainbow Trout Oncorhynchus mykiss

Hematological studies provide essential information about the health of animals, which is crucial for veterinary medicine, scientific research, and aquaculture. Automatic hematological analyzers are an alternative to manual methods, offering faster and more reliable results. The objective of this study was to validate the Sysmex XN-1000V automatic hematology analyzer for blood samples from rainbow trout (<i>Oncorhynchus mykiss</i>), examine the effects of two anticoagulants (K<sub>2</sub>EDTA and lithium heparin), and establish normal blood reference values for this fish species. Additionally, comparative studies were conducted between the Sysmex XN-1000V and manual methods (hemocytometer cell count and blood smear estimation), and reference intervals were established. Ninety-nine heparinized blood samples were analyzed for validation and sample stability tests. The results showed extremely good precision, with a coefficient of variation (CV) below 3% for RBCs, HGB, and HCT and less than 5% for non-RBC cells (leukocytes plus thrombocytes). However, heterophils (%) exhibited higher variability, with a CV of 15.08%. Linearity was excellent, and the carry-over was below 1% for all parameters. The sample stability test indicated that samples could be analyzed for up to 48 h when stored at 4 °C and up to 24 h at room temperature. Non-RBC cells were the first to degrade over time. The automated and manual methods demonstrated good correlation and agreement, validating the analyzer's accuracy. The effects of two anticoagulants, K<sub>2</sub>EDTA and lithium heparin, on the blood samples were also studied. Heparin was the preferred anticoagulant for routine hematological analysis of rainbow trout blood with the Sysmex XN-1000V analyzer. In conclusion, the Sysmex XN-1000V enables complete hemogram analyses to be performed quickly and accurately, standardizing techniques, harmonizing results, and providing reliable reference intervals with <i>O mykiss</i> blood.

Authors

Institutions

Publication Details

Journal
Animals
Published
2025-04-29
DOI
https://doi.org/10.3390/ani15091265
Citations
2
Primary Topic
Fish Ecology and Management Studies
Type
article
Field-Weighted Citation Impact
4.63

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Preliminary Evaluation of an Automated Blood Cell Analyzer for Its Use with Blood Samples from Rainbow Trout Oncorhynchus mykiss

Emmanuel Serrano, Rafaela Cuenca, I. Brandts, Montse Mesalles et al.
2 citations
Animals
Fish Ecology and Management Studies
4.63
article

Preliminary Evaluation of an Automated Blood Cell Analyzer for Its Use with Blood Samples from Rainbow Trout Oncorhynchus mykiss

Emmanuel Serrano, Rafaela Cuenca, I. Brandts, Montse Mesalles, Josep Pastor, Mariana Teles, Meritxell Uroz
article en
2 citations

Abstract

Hematological studies provide essential information about the health of animals, which is crucial for veterinary medicine, scientific research, and aquaculture. Automatic hematological analyzers are an alternative to manual methods, offering faster and more reliable results. The objective of this study was to validate the Sysmex XN-1000V automatic hematology analyzer for blood samples from rainbow trout (<i>Oncorhynchus mykiss</i>), examine the effects of two anticoagulants (K<sub>2</sub>EDTA and lithium heparin), and establish normal blood reference values for this fish species. Additionally, comparative studies were conducted between the Sysmex XN-1000V and manual methods (hemocytometer cell count and blood smear estimation), and reference intervals were established. Ninety-nine heparinized blood samples were analyzed for validation and sample stability tests. The results showed extremely good precision, with a coefficient of variation (CV) below 3% for RBCs, HGB, and HCT and less than 5% for non-RBC cells (leukocytes plus thrombocytes). However, heterophils (%) exhibited higher variability, with a CV of 15.08%. Linearity was excellent, and the carry-over was below 1% for all parameters. The sample stability test indicated that samples could be analyzed for up to 48 h when stored at 4 °C and up to 24 h at room temperature. Non-RBC cells were the first to degrade over time. The automated and manual methods demonstrated good correlation and agreement, validating the analyzer's accuracy. The effects of two anticoagulants, K<sub>2</sub>EDTA and lithium heparin, on the blood samples were also studied. Heparin was the preferred anticoagulant for routine hematological analysis of rainbow trout blood with the Sysmex XN-1000V analyzer. In conclusion, the Sysmex XN-1000V enables complete hemogram analyses to be performed quickly and accurately, standardizing techniques, harmonizing results, and providing reliable reference intervals with <i>O mykiss</i> blood.

AnimalsVol. 15(9)
Universitat Autònoma de Barcelona (ES)
Ministerio de Ciencia e Innovación
Openalex Percentile: Top 7%
Fish Ecology and Management Studies
4.63
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.