M22 – the electron accelerator for calibration and research in radiation dosimetry at METAS

This overview aims to present the metrologically relevant aspects as well as applications beyond conventional radiation dosimetry of the irradiation facility installed at the Swiss Federal Institute of Metrology (METAS). The accelerator, based on the microtron principle, can deliver electron beams to three different beamlines. At the treatment head, the properties of the conventional photon and electron radiation field in water have been characterized in terms of quality specifier TPR 20,10 and R 50 respectively. In a second beamline delivering a narrow electron beam, the focus is on the characterization of the beam energy spectrum using a magnetic spectrometer as well as the beam intensity using an absolutely calibrated integrating current transformer. These two components allow the calibration of Fricke solution, a chemical dosimeter to be used as a primary standard for electron beams. In addition to these two long-established beamlines, a recently installed third beamline delivers a narrow electron beam offering great flexibility for exploration. Its use is not limited to ultra-high dose rates but also, for example, to determining the cross sections of photonuclear reactions.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2025-04-30
DOI
https://doi.org/10.1016/j.nima.2025.170588
Citations
1
Primary Topic
Radiation Effects and Dosimetry
Type
article
Field-Weighted Citation Impact
1.16

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

M22 – the electron accelerator for calibration and research in radiation dosimetry at METAS

Peter Peier, Sándor Vörös, Franziska Frei, Matthias Lüthi
1 citations
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Radiation Effects and Dosimetry
1.16
article

M22 – the electron accelerator for calibration and research in radiation dosimetry at METAS

Peter Peier, Sándor Vörös, Franziska Frei, Matthias Lüthi
article en
1 citations

Abstract

This overview aims to present the metrologically relevant aspects as well as applications beyond conventional radiation dosimetry of the irradiation facility installed at the Swiss Federal Institute of Metrology (METAS). The accelerator, based on the microtron principle, can deliver electron beams to three different beamlines. At the treatment head, the properties of the conventional photon and electron radiation field in water have been characterized in terms of quality specifier TPR 20,10 and R 50 respectively. In a second beamline delivering a narrow electron beam, the focus is on the characterization of the beam energy spectrum using a magnetic spectrometer as well as the beam intensity using an absolutely calibrated integrating current transformer. These two components allow the calibration of Fricke solution, a chemical dosimeter to be used as a primary standard for electron beams. In addition to these two long-established beamlines, a recently installed third beamline delivers a narrow electron beam offering great flexibility for exploration. Its use is not limited to ultra-high dose rates but also, for example, to determining the cross sections of photonuclear reactions.

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1077
Swiss Federal Institute of Metrology (CH)
Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
Affordable and clean energy
Openalex Percentile: Top 22%
Radiation Effects and Dosimetry
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.