Dose optimization of cone beam computed tomography: measurement of parotid absorbed dose and image quality assessment

Cone-beam computed tomography (CBCT) is a three-dimensional (3D) imaging technology used in dentistry. This study aimed to reduce radiation exposure by adjusting CBCT parameters while ensuring that image quality remained suitable for diagnostic purposes. This controlled experimental study was conducted between February and July 2022 at the private Alpha Center for Dental Radiology in Erbil, Kurdistan Region of Iraq. The PaX-i3D SMART CBCT system, thermoluminescent dosimeters, and a specialized head and neck dosimetry phantom were used to measure the parotid gland. Tube voltage varied from 75 to 99 kVp, and tube current varied from 4 to 10 mA. For the image quality assessment, a dry human mandible immersed in water was exposed to CBCT X-rays with the same dosimetry exposure settings. Subjective image assessment was performed randomly by four dental and maxillofacial radiologists. The highest absorbed dose recorded was 654.47 µGy at 94 kVp and 8.1 mA, while the lowest was 198.5 µGy at 80 kVp and 4 mA. Out of the 32 scans, 19 images were considered acceptable based on clinical evaluation, and their absorbed dose ranges were lower than the default exposure setting of the device. Statistically, there was a strong positive correlation between absorbed dose, kVp, and mA, and a non-significant correlation between image quality and kVp in five (4, 4.5, 6, 8, and 10) of the seven mA groups. Optimizing CBCT settings to 80 kVp and 4 mA significantly reduced the radiation dose to the parotid gland while maintaining diagnostic image quality. This finding supports the adoption of lower mA and kVp settings in clinical practice to enhance patient safety without compromising diagnostic effectiveness.

Authors

Institutions

Publication Details

Journal
Journal of Medicine and Life
Published
2025-03-01
DOI
https://doi.org/10.25122/jml-2024-0168
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dose optimization of cone beam computed tomography: measurement of parotid absorbed dose and image quality assessment

Shereen Ismail Hajee, Lezan Othman Shina
Journal of Medicine and Life
Dental Radiography and Imaging
article

Dose optimization of cone beam computed tomography: measurement of parotid absorbed dose and image quality assessment

Shereen Ismail Hajee, Lezan Othman Shina
article en

Abstract

Cone-beam computed tomography (CBCT) is a three-dimensional (3D) imaging technology used in dentistry. This study aimed to reduce radiation exposure by adjusting CBCT parameters while ensuring that image quality remained suitable for diagnostic purposes. This controlled experimental study was conducted between February and July 2022 at the private Alpha Center for Dental Radiology in Erbil, Kurdistan Region of Iraq. The PaX-i3D SMART CBCT system, thermoluminescent dosimeters, and a specialized head and neck dosimetry phantom were used to measure the parotid gland. Tube voltage varied from 75 to 99 kVp, and tube current varied from 4 to 10 mA. For the image quality assessment, a dry human mandible immersed in water was exposed to CBCT X-rays with the same dosimetry exposure settings. Subjective image assessment was performed randomly by four dental and maxillofacial radiologists. The highest absorbed dose recorded was 654.47 µGy at 94 kVp and 8.1 mA, while the lowest was 198.5 µGy at 80 kVp and 4 mA. Out of the 32 scans, 19 images were considered acceptable based on clinical evaluation, and their absorbed dose ranges were lower than the default exposure setting of the device. Statistically, there was a strong positive correlation between absorbed dose, kVp, and mA, and a non-significant correlation between image quality and kVp in five (4, 4.5, 6, 8, and 10) of the seven mA groups. Optimizing CBCT settings to 80 kVp and 4 mA significantly reduced the radiation dose to the parotid gland while maintaining diagnostic image quality. This finding supports the adoption of lower mA and kVp settings in clinical practice to enhance patient safety without compromising diagnostic effectiveness.

Journal of Medicine and LifeVol. 18(3)
Hawler Medical University (IQ), University of Kurdistan Hewler (IQ)
Clean water and sanitation
Openalex Percentile: Top 88%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.