Dose calculation in nuclear medicine with magnetic resonance imaging images using Monte Carlo method
In recent years, scientists have been trying to convert magnetic resonance imaging (MRI) images into computed tomography (CT) images for dose calculations while taking advantage of the benefits of MRI images. The main approaches for image conversion are bulk density, Atlas registration, and machine learning. These methods have limitations in accuracy and time consumption and require large datasets to convert images. In this study, the novel 'voxels spawn voxels' technique combined with the 'orthonormalize' feature in Carimas software was developed to build a conversion dataset from MRI intensity to Hounsfield unit value for some structural regions including gluteus maximus, liver, kidneys, spleen, pancreas, and colon. The original CT images and the converted MRI images were imported into the Geant4/Gamos software for dose calculation. It gives good results (<5%) in most organs except the intestine (18%).
Authors
- Nguyễn Thị Phương Thảo (ORCID: https://orcid.org/0000-0003-0150-1774)
- Nguyễn Thiên Trung
- Trương Thị Hồng Loan (ORCID: https://orcid.org/0000-0001-9625-9954)
- Pham Vo Trung Hau
- Le Huy Vu
Institutions
- Pham Ngoc Thach University of Medicine (VN)
- Ho Chi Minh City University of Science (VN)
- Ho Chi Minh City University of Education (VN)
Publication Details
- Journal
- Radiation Protection Dosimetry
- Published
- 2025-04-11
- DOI
- https://doi.org/10.1093/rpd/ncaf043
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00