Validation of an Automated Cell Counter Method for HLA-DR and CD3 Expression in Cells Obtained from Low Volume Human Tears

<b>Background/Objectives</b>: Tears are a promising source of biomarkers reflecting both ocular and systemic conditions. However, small sample volumes and low cell yields pose technical challenges in analytical workflows. This study aimed to evaluate the feasibility of quantifying total cell counts and characterizing HLA-DR and CD3 expression in tear-derived cells using an automated cell counter with fluorescence detection (Countess 3 FL). <b>Methods</b>: Tears were collected from 31 patients, centrifuged and the resulting pellet was incubated with HLA-DR and CD3 antibodies, markers of inflammation and T lymphocytes, respectively. Data obtained from Countess 3 FL were compared with conventional flow cytometry and immunofluorescence. For technical performance analysis, precision and reproducibility of cell count and staining were measured. For method validation, an in vitro model of hyperosmolar stress was assessed by culturing conjunctival epithelial cells (CCL20.2) with 350 or 450 mOsm NaCl. <b>Results</b>: The total cell yield in each tear sample correlated with the tear surnatant volume, in a range of 1-40μL (mean total cell number: 1.3 ± 1.1 × 10<sup>4</sup>, correlation analysis with tear volume: r = 0.47, <i>p</i> < 0.05). HLA-DR and CD3 were detected in all samples, with a mean value, respectively, of 43.6% (±21.0) and 25.0% (±15.0) intensity. Data were comparable to those obtained from standard flow cytometry analysis.HLA-DR increase in CCL20.2 exposed to hyperosmolar stress was recorded using Countess 3FL reading, confirming the detection capacity of the proposed method. <b>Conclusions</b>: The automated cell counter can provide HLA-DR and CD3 quantification in tear cell samples, despite the high variability and the low volume availability of tear samples. Method standardization and technical improvements are necessary to strengthen this application in the clinical setting.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-04-28
DOI
https://doi.org/10.3390/diagnostics15091124
Citations
1
Primary Topic
Ocular Surface and Contact Lens
Type
article
Field-Weighted Citation Impact
1.88
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of an Automated Cell Counter Method for HLA-DR and CD3 Expression in Cells Obtained from Low Volume Human Tears

Carmen Ciavarella, Gianandrea Pasquinelli, Michele Potenza, Piera Versura et al.
1 citations
Diagnostics
Ocular Surface and Contact Lens
1.88
article

Validation of an Automated Cell Counter Method for HLA-DR and CD3 Expression in Cells Obtained from Low Volume Human Tears

Carmen Ciavarella, Gianandrea Pasquinelli, Michele Potenza, Piera Versura, Annalisa Astolfi, Chiara Coslovi, Luigi Fontana
article en
1 citations

Abstract

<b>Background/Objectives</b>: Tears are a promising source of biomarkers reflecting both ocular and systemic conditions. However, small sample volumes and low cell yields pose technical challenges in analytical workflows. This study aimed to evaluate the feasibility of quantifying total cell counts and characterizing HLA-DR and CD3 expression in tear-derived cells using an automated cell counter with fluorescence detection (Countess 3 FL). <b>Methods</b>: Tears were collected from 31 patients, centrifuged and the resulting pellet was incubated with HLA-DR and CD3 antibodies, markers of inflammation and T lymphocytes, respectively. Data obtained from Countess 3 FL were compared with conventional flow cytometry and immunofluorescence. For technical performance analysis, precision and reproducibility of cell count and staining were measured. For method validation, an in vitro model of hyperosmolar stress was assessed by culturing conjunctival epithelial cells (CCL20.2) with 350 or 450 mOsm NaCl. <b>Results</b>: The total cell yield in each tear sample correlated with the tear surnatant volume, in a range of 1-40μL (mean total cell number: 1.3 ± 1.1 × 10<sup>4</sup>, correlation analysis with tear volume: r = 0.47, <i>p</i> < 0.05). HLA-DR and CD3 were detected in all samples, with a mean value, respectively, of 43.6% (±21.0) and 25.0% (±15.0) intensity. Data were comparable to those obtained from standard flow cytometry analysis.HLA-DR increase in CCL20.2 exposed to hyperosmolar stress was recorded using Countess 3FL reading, confirming the detection capacity of the proposed method. <b>Conclusions</b>: The automated cell counter can provide HLA-DR and CD3 quantification in tear cell samples, despite the high variability and the low volume availability of tear samples. Method standardization and technical improvements are necessary to strengthen this application in the clinical setting.

DiagnosticsVol. 15(9)
Azienda USL di Bologna (IT), University of Bologna (IT)
Openalex Percentile: Top 16%
Ocular Surface and Contact Lens
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.