Design of a high-resolution liquid xenon detector for positron emission tomography

Positron Emission Tomography (PET) is a vital imaging technique extensively used for early cancer detection by visualizing metabolic processes in the body. While traditional PET systems use scintillation crystals like bismuth germanate (BGO) or lutetium oxyorthosilicate (LSO) to detect gamma rays, they have inherent energy and spatial resolution limitations. This paper proposes an advanced PET design using liquid xenon (LXe)-based detectors that integrate scintillation and ionization energy detection. Our PET detector design has a monolithic liquid xenon target of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="m1"><mml:mrow><mml:mn>5</mml:mn><mml:mo>×</mml:mo><mml:mn>5</mml:mn><mml:mo>×</mml:mo><mml:mn>5</mml:mn><mml:mtext> </mml:mtext><mml:msup><mml:mrow><mml:mi mathvariant="normal">c</mml:mi><mml:mi mathvariant="normal">m</mml:mi></mml:mrow><mml:mrow><mml:mn>3</mml:mn></mml:mrow></mml:msup></mml:mrow></mml:math> , from where scintillation light is detected by silicon photomultipliers (SiPMs) placed on one side of the target. The ionization is converted to field-enhanced electroluminescence in liquid xenon and detected by the same SiPMs. We use Monte Carlo simulations to optimize the configuration of the electric field and improve the light collection efficiency. Combining both detection modes, the proposed system aims to significantly improve the energy resolution to approximately 2% full width at half maximum (FWHM). Furthermore, machine learning models enhance position reconstruction accuracy with sub-millimeter horizontal and depth-of-interaction (DOI) resolutions. The results indicate that the LXe-based PET detector can achieve superior performance compared to current PET technologies, offering enhanced imaging accuracy with the potential for reduced doses of radioactive tracer.

Authors

Institutions

Publication Details

Journal
Frontiers in Detector Science and Technology
Published
2025-04-28
DOI
https://doi.org/10.3389/fdest.2025.1488822
Citations
1
Primary Topic
Atomic and Subatomic Physics Research
Type
article
Field-Weighted Citation Impact
1.25
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design of a high-resolution liquid xenon detector for positron emission tomography

J. H. Feng, Min Zhong, M. J. Ni, Alexander Backues
1 citations
Frontiers in Detector Science and Technology
Atomic and Subatomic Physics Research
1.25
article

Design of a high-resolution liquid xenon detector for positron emission tomography

J. H. Feng, Min Zhong, M. J. Ni, Alexander Backues
article en
1 citations

Abstract

Positron Emission Tomography (PET) is a vital imaging technique extensively used for early cancer detection by visualizing metabolic processes in the body. While traditional PET systems use scintillation crystals like bismuth germanate (BGO) or lutetium oxyorthosilicate (LSO) to detect gamma rays, they have inherent energy and spatial resolution limitations. This paper proposes an advanced PET design using liquid xenon (LXe)-based detectors that integrate scintillation and ionization energy detection. Our PET detector design has a monolithic liquid xenon target of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="m1"><mml:mrow><mml:mn>5</mml:mn><mml:mo>×</mml:mo><mml:mn>5</mml:mn><mml:mo>×</mml:mo><mml:mn>5</mml:mn><mml:mtext> </mml:mtext><mml:msup><mml:mrow><mml:mi mathvariant="normal">c</mml:mi><mml:mi mathvariant="normal">m</mml:mi></mml:mrow><mml:mrow><mml:mn>3</mml:mn></mml:mrow></mml:msup></mml:mrow></mml:math> , from where scintillation light is detected by silicon photomultipliers (SiPMs) placed on one side of the target. The ionization is converted to field-enhanced electroluminescence in liquid xenon and detected by the same SiPMs. We use Monte Carlo simulations to optimize the configuration of the electric field and improve the light collection efficiency. Combining both detection modes, the proposed system aims to significantly improve the energy resolution to approximately 2% full width at half maximum (FWHM). Furthermore, machine learning models enhance position reconstruction accuracy with sub-millimeter horizontal and depth-of-interaction (DOI) resolutions. The results indicate that the LXe-based PET detector can achieve superior performance compared to current PET technologies, offering enhanced imaging accuracy with the potential for reduced doses of radioactive tracer.

Frontiers in Detector Science and TechnologyVol. 3
Bishop's University (CA), University of California San Diego (US)
Openalex Percentile: Top 22%
Atomic and Subatomic Physics Research
1.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.