Endpoint attachment of unfractionated vs low molecular weight heparin: a comparative study on blood compatibility

Introduction Blood compatibility is a critical requirement for materials used in medical devices to prevent adverse reactions such as thrombosis and inflammation. Heparin, a natural anticoagulant, is commonly used to coat medical devices such as catheters and stents, mimicking the endothelial glycocalyx and thereby preventing thrombus formation. While endpoint attached (EPA) heparin surfaces have proven effective in clinical use, concerns have arisen regarding heparin-induced thrombocytopenia (HIT) associated with unfractionated heparin (UFH)-bonded prostheses. Given the lower HIT risk profile of low molecular weight heparin (LMWH), this study aimed to evaluate the blood compatibility of LMWH-functionalized surfaces in comparison to those functionalized with UFH. Methods Conventional 96-well plates and adsorbent materials were functionalized with either UFH or LMWH. The blood compatibility of these surfaces was assessed by analyzing coagulation parameters, platelet adhesion, and leukocyte adhesion. Antithrombin III (AT-III) binding was also examined to understand the anticoagulant mechanism of LMWH-functionalized surfaces. Results LMWH-functionalized surfaces demonstrated no significant binding to AT-III. Despite this, both LMWH- and UFH-functionalized surfaces effectively prevented the activation of coagulation triggered by foreign surfaces. Furthermore, there was no observed increase in platelet or leukocyte adhesion on LMWH-functionalized surfaces compared to UFH-functionalized ones, indicating a similar degree of blood compatibility between the two. Discussion This study highlights that LMWH can provide blood compatibility comparable to that of UFH when used for surface functionalization, despite its lack of AT-III binding. These findings support the potential of LMWH as a safer alternative to UFH in medical device coatings, particularly in applications where the risk of HIT must be minimized.

Authors

Institutions

Publication Details

Journal
Frontiers in Materials
Published
2025-04-28
DOI
https://doi.org/10.3389/fmats.2025.1557939
Primary Topic
Heparin-Induced Thrombocytopenia and Thrombosis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Endpoint attachment of unfractionated vs low molecular weight heparin: a comparative study on blood compatibility

Jennifer Zottl, Christoph Bauer, Viktoria Weber, Claudia Schildböck et al.
1 links
Frontiers in Materials
Heparin-Induced Thrombocytopenia and Thrombosis
article

Endpoint attachment of unfractionated vs low molecular weight heparin: a comparative study on blood compatibility

Jennifer Zottl, Christoph Bauer, Viktoria Weber, Claudia Schildböck, Stephan Harm, Denisa Cont
article en

Abstract

Introduction Blood compatibility is a critical requirement for materials used in medical devices to prevent adverse reactions such as thrombosis and inflammation. Heparin, a natural anticoagulant, is commonly used to coat medical devices such as catheters and stents, mimicking the endothelial glycocalyx and thereby preventing thrombus formation. While endpoint attached (EPA) heparin surfaces have proven effective in clinical use, concerns have arisen regarding heparin-induced thrombocytopenia (HIT) associated with unfractionated heparin (UFH)-bonded prostheses. Given the lower HIT risk profile of low molecular weight heparin (LMWH), this study aimed to evaluate the blood compatibility of LMWH-functionalized surfaces in comparison to those functionalized with UFH. Methods Conventional 96-well plates and adsorbent materials were functionalized with either UFH or LMWH. The blood compatibility of these surfaces was assessed by analyzing coagulation parameters, platelet adhesion, and leukocyte adhesion. Antithrombin III (AT-III) binding was also examined to understand the anticoagulant mechanism of LMWH-functionalized surfaces. Results LMWH-functionalized surfaces demonstrated no significant binding to AT-III. Despite this, both LMWH- and UFH-functionalized surfaces effectively prevented the activation of coagulation triggered by foreign surfaces. Furthermore, there was no observed increase in platelet or leukocyte adhesion on LMWH-functionalized surfaces compared to UFH-functionalized ones, indicating a similar degree of blood compatibility between the two. Discussion This study highlights that LMWH can provide blood compatibility comparable to that of UFH when used for surface functionalization, despite its lack of AT-III binding. These findings support the potential of LMWH as a safer alternative to UFH in medical device coatings, particularly in applications where the risk of HIT must be minimized.

Frontiers in MaterialsVol. 12
Universität für Weiterbildung Krems (AT), Karl Landsteiner University of Health Sciences (AT)
Openalex Percentile: Top 85%
Heparin-Induced Thrombocytopenia and Thrombosis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.