Thermal dose feedback control systems applied to magnetic nanoparticle hyperthermia

Clinical magnetic nanoparticle hyperthermia therapy (MNHT) requires controlled energy deposition to achieve a prescribed tumor thermal dose. The objective of this work is to design a thermal dose feedback control to deliver prescribed Cumulative Equivalent Minutes at 43 [°C] (CEM43) based on values at selected tumor boundary points. Constraints were imposed to maintain the maximum treatment temperature below 60 [°C] and the tumor boundary at ∼ 43 [°C]. The controller was designed by performing an integrated system dynamic - finite element analysis. Finite element-bioheat transfer (FE-BHT) simulations were performed on a computational phantom developed from the imaging data of a de-identified human head divided into voxels representing the skull, cerebrospinal fluid (CSF), brain, tumor, and ventricles. A uniform distribution of magnetic nanoparticles (MNPs) in an ellipsoid was used to represent MNPs in the phantom tumor. The MNP distribution was subdivided into three domains to simulate the steerable spatially confined heating region during MNHT. Proportional-integral-derivative (PID) control and model predictive control (MPC) were explored. Regions of the phantom tumor that were undertreated during the simulated MNHT were selectively heated by adjusting the heating volume to improve the tumor coverage index (CI; tumor volume ≥ CEM43 of 20 [min]). Results show that steerable spatially confined heating improves CI by ∼15%. MPC achieves CI of 80% faster than PID (67 [min] vs. 80 [min]). Simulations demonstrated the feasibility of automated control to deliver tumor conformal thermal doses using steerable spatially confined heating.

Authors

Institutions

Publication Details

Journal
International Journal of Hyperthermia
Published
2025-04-27
DOI
https://doi.org/10.1080/02656736.2025.2491519
Citations
4
Primary Topic
Characterization and Applications of Magnetic Nanoparticles
Type
article
Field-Weighted Citation Impact
2.70
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Thermal dose feedback control systems applied to magnetic nanoparticle hyperthermia

Robert Ivkov, Ma’moun Abu-Ayyad, Nageshwar Arepally, Costas G. Hadjipanayis et al.
4 citations
International Journal of Hyperthermia
Characterization and Applications of Magnetic Nanoparticles
2.70
article

Thermal dose feedback control systems applied to magnetic nanoparticle hyperthermia

Robert Ivkov, Ma’moun Abu-Ayyad, Nageshwar Arepally, Costas G. Hadjipanayis, Hayden Carlton, Anilchandra Attaluri, Yash Sharad Lad, Shreeniket Pawar
article en
4 citations

Abstract

Clinical magnetic nanoparticle hyperthermia therapy (MNHT) requires controlled energy deposition to achieve a prescribed tumor thermal dose. The objective of this work is to design a thermal dose feedback control to deliver prescribed Cumulative Equivalent Minutes at 43 [°C] (CEM43) based on values at selected tumor boundary points. Constraints were imposed to maintain the maximum treatment temperature below 60 [°C] and the tumor boundary at ∼ 43 [°C]. The controller was designed by performing an integrated system dynamic - finite element analysis. Finite element-bioheat transfer (FE-BHT) simulations were performed on a computational phantom developed from the imaging data of a de-identified human head divided into voxels representing the skull, cerebrospinal fluid (CSF), brain, tumor, and ventricles. A uniform distribution of magnetic nanoparticles (MNPs) in an ellipsoid was used to represent MNPs in the phantom tumor. The MNP distribution was subdivided into three domains to simulate the steerable spatially confined heating region during MNHT. Proportional-integral-derivative (PID) control and model predictive control (MPC) were explored. Regions of the phantom tumor that were undertreated during the simulated MNHT were selectively heated by adjusting the heating volume to improve the tumor coverage index (CI; tumor volume ≥ CEM43 of 20 [min]). Results show that steerable spatially confined heating improves CI by ∼15%. MPC achieves CI of 80% faster than PID (67 [min] vs. 80 [min]). Simulations demonstrated the feasibility of automated control to deliver tumor conformal thermal doses using steerable spatially confined heating.

International Journal of HyperthermiaVol. 42(1)
Pennsylvania State University (US), Johns Hopkins University (US), University of Pittsburgh (US), Johns Hopkins Medicine (US), Division of Materials Science and Engineering (AU), Neurological Surgery (US), Sidney Kimmel Comprehensive Cancer Center (US)
Openalex Percentile: Top 11%
Characterization and Applications of Magnetic Nanoparticles
2.70
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.