Simulation in Transcutaneous Injection Laryngoplasty Using a 3D-printed Laryngeal Injection Task Trainer (LITT)

Objective: To develop a low-cost 3D-printed Laryngeal Injection Task Trainer (LITT) for simulation of transcutaneous injection laryngoplasty and evaluate its effectiveness among trainees. Methods:Design: Cross-Sectional Time-Motion Study and Post-Training Evaluation Survey Setting: Tertiary University College of Medicine Anatomy LaboratoryParticipants: Five residents and 2 consultants in Otolaryngology-Head and Neck Surgery were trained using a LITT fabricated through three-dimensional printing of cartilage models and silicone molds through the UP SIBOL Greenhouse and the Philippine General Hospital Craniomaxillofacial Prosthesis Unit. Participants were asked to perform transcutaneous laryngeal injection on the LITT. The duration in performing the procedure from palpation of landmarks to injection of material was measured in seconds and recorded. After the simulation workshop, participants were asked to evaluate the subjective experience with surgical simulation using the modified Michigan Standard Simulation Experience Scale (MiSSES). Results: The mean time in performing transcutaneous laryngeal injection across different training levels for either trans-cricothyroid or trans-thyrohyoid approach was not statistically significant. Mean time in performing transcutaneous laryngeal injection between 2 approaches, the trans-cricothyroid approach (79.32 seconds with SD 34.89) and the trans-thyrohyoid approach (35.16 seconds with SD of 10.88), was statistically significant. Using the MiSSES tool for evaluation of the simulation training, majority of the participants agreed that the LITT allowed self-efficacy in performing the procedure, had adequate fidelity in terms of its characteristics and components, had high educational value in training participants for knowledge and skills on transcutaneous laryngeal injection and that the simulation provided good teaching quality. For the overall rating, all participants concurred, with 83.3% of participants answering strongly agree and 16.7% of participants responding somewhat agree, that the simulation experience was helpful in knowledge and skills development on transcutaneous laryngeal injection. On final evaluation of the LITT, 75% of the participants responded that it can be used in training but should be improved slightly. Conclusion: The laryngeal injection task trainer (LITT) had an acceptable fidelity, educational value, teaching quality to improve self-efficacy in performing transcutaneous injection laryngoplasty - with points for improvement in terms of model and set-up stability, color accuracy and consistency of the visual indicator. It is recommended that the LITT be also used for future workshops and possibly further validation studies after applying the points for improvement gathered from this simulation study.

Authors

Publication Details

Journal
Philippine Journal of Otolaryngology Head and Neck Surgery
Published
2025-04-27
DOI
https://doi.org/10.32412/pjohns.v40i1.2369
Primary Topic
Voice and Speech Disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simulation in Transcutaneous Injection Laryngoplasty Using a 3D-printed Laryngeal Injection Task Trainer (LITT)

Ryner Jose D. Carrillo, Miguel Aljibe, Philip Fullante, Anna Pamela Dela Cruz
1 links
Philippine Journal of Otolaryngology Head and Neck Surgery
Voice and Speech Disorders
article

Simulation in Transcutaneous Injection Laryngoplasty Using a 3D-printed Laryngeal Injection Task Trainer (LITT)

Ryner Jose D. Carrillo, Miguel Aljibe, Philip Fullante, Anna Pamela Dela Cruz
article en

Abstract

Objective: To develop a low-cost 3D-printed Laryngeal Injection Task Trainer (LITT) for simulation of transcutaneous injection laryngoplasty and evaluate its effectiveness among trainees. Methods:Design: Cross-Sectional Time-Motion Study and Post-Training Evaluation Survey Setting: Tertiary University College of Medicine Anatomy LaboratoryParticipants: Five residents and 2 consultants in Otolaryngology-Head and Neck Surgery were trained using a LITT fabricated through three-dimensional printing of cartilage models and silicone molds through the UP SIBOL Greenhouse and the Philippine General Hospital Craniomaxillofacial Prosthesis Unit. Participants were asked to perform transcutaneous laryngeal injection on the LITT. The duration in performing the procedure from palpation of landmarks to injection of material was measured in seconds and recorded. After the simulation workshop, participants were asked to evaluate the subjective experience with surgical simulation using the modified Michigan Standard Simulation Experience Scale (MiSSES). Results: The mean time in performing transcutaneous laryngeal injection across different training levels for either trans-cricothyroid or trans-thyrohyoid approach was not statistically significant. Mean time in performing transcutaneous laryngeal injection between 2 approaches, the trans-cricothyroid approach (79.32 seconds with SD 34.89) and the trans-thyrohyoid approach (35.16 seconds with SD of 10.88), was statistically significant. Using the MiSSES tool for evaluation of the simulation training, majority of the participants agreed that the LITT allowed self-efficacy in performing the procedure, had adequate fidelity in terms of its characteristics and components, had high educational value in training participants for knowledge and skills on transcutaneous laryngeal injection and that the simulation provided good teaching quality. For the overall rating, all participants concurred, with 83.3% of participants answering strongly agree and 16.7% of participants responding somewhat agree, that the simulation experience was helpful in knowledge and skills development on transcutaneous laryngeal injection. On final evaluation of the LITT, 75% of the participants responded that it can be used in training but should be improved slightly. Conclusion: The laryngeal injection task trainer (LITT) had an acceptable fidelity, educational value, teaching quality to improve self-efficacy in performing transcutaneous injection laryngoplasty - with points for improvement in terms of model and set-up stability, color accuracy and consistency of the visual indicator. It is recommended that the LITT be also used for future workshops and possibly further validation studies after applying the points for improvement gathered from this simulation study.

Philippine Journal of Otolaryngology Head and Neck SurgeryVol. 40(1)
University of the Philippines
Openalex Percentile: Top 89%
Voice and Speech Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.