Periodic blinking manipulation of magnetic Janus particles with a tunable electromagnetic field for rapid sensing of extracellular vesicles

Detecting small biological molecules is challenging due to their tiny size, vulnerability, and low concentrations in samples. Bead-based biosensors are frequently used as probes but require tedious processing or expensive instruments. By combining magnetic Janus particles (MJPs) and an electromagnetic device, we successfully built an active diagnostic tool for the rapid sensing of small extracellular vesicles (sEVs). We observed that the system can be altered according to particle size, distance between MJPs and the electromagnet, fluid viscosity, and magnetic field strength. By modulating the driving frequency from low (3 Hz) to high (22 Hz), the MJPs gradually lose their synchrony with the external magnetic field after exceeding a certain threshold termed cutoff frequency. The novel sEVs sensing MJP system was characterized through both theoretical and experimental methods, showing reliable performance in identifying the cancer cell OECM-1-derived sEVs using the CD63 surface marker. A decent sEV concentration of 2.9 × 10<sup>9</sup> particles mL<sup>-1</sup> was reached and a high specificity was also observed. This approach opens a door for the realization of disease screening, such as cancer, using intact exosomes from body fluids without sophisticated processing. These findings provide insight into the future use of MJPs as point-of-care testing tools for liquid biopsy.

Authors

Institutions

Publication Details

Journal
Frontiers in Bioengineering and Biotechnology
Published
2025-04-25
DOI
https://doi.org/10.3389/fbioe.2025.1565479
Citations
1
Primary Topic
Characterization and Applications of Magnetic Nanoparticles
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Periodic blinking manipulation of magnetic Janus particles with a tunable electromagnetic field for rapid sensing of extracellular vesicles

Chi‐Ying F. Huang, Jhih-Cheng Wang, Thi Thanh Huong Pham, Yu-Hsuan Chou et al.
1 citations
Frontiers in Bioengineering and Biotechnology
Characterization and Applications of Magnetic Nanoparticles
0.67
article

Periodic blinking manipulation of magnetic Janus particles with a tunable electromagnetic field for rapid sensing of extracellular vesicles

Chi‐Ying F. Huang, Jhih-Cheng Wang, Thi Thanh Huong Pham, Yu-Hsuan Chou, Tai‐Hua Yang, Ting‐Yuan Tu, Han-Sheng Chuang
article en
1 citations

Abstract

Detecting small biological molecules is challenging due to their tiny size, vulnerability, and low concentrations in samples. Bead-based biosensors are frequently used as probes but require tedious processing or expensive instruments. By combining magnetic Janus particles (MJPs) and an electromagnetic device, we successfully built an active diagnostic tool for the rapid sensing of small extracellular vesicles (sEVs). We observed that the system can be altered according to particle size, distance between MJPs and the electromagnet, fluid viscosity, and magnetic field strength. By modulating the driving frequency from low (3 Hz) to high (22 Hz), the MJPs gradually lose their synchrony with the external magnetic field after exceeding a certain threshold termed cutoff frequency. The novel sEVs sensing MJP system was characterized through both theoretical and experimental methods, showing reliable performance in identifying the cancer cell OECM-1-derived sEVs using the CD63 surface marker. A decent sEV concentration of 2.9 × 10<sup>9</sup> particles mL<sup>-1</sup> was reached and a high specificity was also observed. This approach opens a door for the realization of disease screening, such as cancer, using intact exosomes from body fluids without sophisticated processing. These findings provide insight into the future use of MJPs as point-of-care testing tools for liquid biopsy.

Frontiers in Bioengineering and BiotechnologyVol. 13
National Sun Yat-sen University (TW), National Yang Ming Chiao Tung University (TW), Chi Mei Medical Center (TW), National Cheng Kung University Hospital (TW), National Cheng Kung University (TW), Southern Taiwan University of Science and Technology (TW)
National Science and Technology Council
Openalex Percentile: Top 32%
Characterization and Applications of Magnetic Nanoparticles
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.