Measurement of U and Pu isotopic ratio in environmental samples through MC-ICP-MS integrated to an automated microfluidic column

Accurate identification of isotopes and elements of uranium (U) and plutonium (Pu) in environmental samples from the nuclear fuel cycle is essential for environmental monitoring and safeguarding. A novel microfluidic system coupled with a multi-collecting inductively coupled plasma mass spectrometer (MC-ICP-MS) allows the measurement Pu and U isotope ratios in real time. The application reduces the sample volume required to less than 80 µL with required quantities of analytes being reduced from the conventional micrograms to less than nanograms. This reduction in sample and reagent volume makes the analytical process less resource-intensive and generates less waste; a significant benefit for safer and more efficient analysis operations.

Authors

Institutions

Publication Details

Journal
Journal of Radioanalytical and Nuclear Chemistry
Published
2025-04-24
DOI
https://doi.org/10.1007/s10967-025-10097-z
Primary Topic
Radioactive contamination and transfer
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measurement of U and Pu isotopic ratio in environmental samples through MC-ICP-MS integrated to an automated microfluidic column

Matthew Higginson, S.L. Heath, Bernard J. Treves Brown, Shuang Yu Han et al.
Journal of Radioanalytical and Nuclear Chemistry
Radioactive contamination and transfer
article

Measurement of U and Pu isotopic ratio in environmental samples through MC-ICP-MS integrated to an automated microfluidic column

Matthew Higginson, S.L. Heath, Bernard J. Treves Brown, Shuang Yu Han, Philip Kaye, Sidney R. Hemming, Clint A. Sharrad
article en

Abstract

Accurate identification of isotopes and elements of uranium (U) and plutonium (Pu) in environmental samples from the nuclear fuel cycle is essential for environmental monitoring and safeguarding. A novel microfluidic system coupled with a multi-collecting inductively coupled plasma mass spectrometer (MC-ICP-MS) allows the measurement Pu and U isotope ratios in real time. The application reduces the sample volume required to less than 80 µL with required quantities of analytes being reduced from the conventional micrograms to less than nanograms. This reduction in sample and reagent volume makes the analytical process less resource-intensive and generates less waste; a significant benefit for safer and more efficient analysis operations.

Journal of Radioanalytical and Nuclear ChemistryVol. 334(5)
National Oceanography Centre (GB), AWE Nuclear Security Technologies (GB), University of Manchester (GB), Space Research Institute (UA)
University of Southampton, Engineering and Physical Sciences Research Council
Responsible consumption and production
Openalex Percentile: Top 91%
Radioactive contamination and transfer
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.