[A clinical study of electrocochleography monitoring for residual hearing retention during minimally invasive cochlear implant].

<b>Objective:</b>To investigate the application value of intraoperative electrocochleography(ECochG) monitoring technique and insertion techniques in cochlear implant(CI) and analyze its relationship with postoperative residual hearing(RH) preservation. <b>Methods:</b>Thirty-one patients(35 ears) who received CI in our hospital from June 2022 to July 2024 were enrolled. The Advanced Bionics Active Insertion Monitoring(AIM) system was used for real-time ECochG monitoring during surgery. Intraoperative cochlear microphonics (CM) waveform changes were recorded and analyzed in relation to postoperative RH preservation. <b>Results:</b>①ECochG recordings were successfully obtained in 34 of 35 ears (97.1%). ②According to Harris classification, there were 7 ears(20.6%) of Type A(rising), 7 ears(20.6%) of Type C(declining), 8 ears(23.5%) of Type CC(fluctuating), and 12 ears(35.3%) of Type D(no response). ③The total CM amplitude decrease was significantly moderately correlated with postoperative low-mid frequency hearing loss(<i>r</i>=0.67, <i>P</i>=0.017). The total CM amplitude decrease was significantly moderately correlated with postoperative low frequency hearing loss(<i>r</i>=0.65, <i>P</i>=0.023). ④For the mean amplitude variation, the Amax was 30.70 μV, the Amin was 8.64 μV, and the Aend was 18.27 μV. ⑤Sixteen cases completed postoperative follow-up, with an average low-mid frequency(125-1 000 Hz) residual hearing loss of 15.25 dB HL and a RH preservation rate of 87.5%. <b>Conclusion:</b>Intraoperative ECochG monitoring can effectively predict postoperative residual hearing changes, effectively guide surgical manipulation, and improve residual hearing preservation rate.

Authors

Institutions

Publication Details

Journal
PubMed
Published
2025-05-01
DOI
https://doi.org/10.13201/j.issn.2096-7993.2025.05.006
Citations
1
Primary Topic
Hearing Loss and Rehabilitation
Type
article
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

[A clinical study of electrocochleography monitoring for residual hearing retention during minimally invasive cochlear implant].

Qinglei Dai, Fangxia Hu, Ye Cao, Yuan Ni et al.
1 citations
PubMed
Hearing Loss and Rehabilitation
article

[A clinical study of electrocochleography monitoring for residual hearing retention during minimally invasive cochlear implant].

Qinglei Dai, Fangxia Hu, Ye Cao, Yuan Ni, Jianfen Luo, Xiuhua Chao, Haibo Wang, Lei Xu, Xiaohui Zhou, Ruijie Wang
article en
1 citations

Abstract

<b>Objective:</b>To investigate the application value of intraoperative electrocochleography(ECochG) monitoring technique and insertion techniques in cochlear implant(CI) and analyze its relationship with postoperative residual hearing(RH) preservation. <b>Methods:</b>Thirty-one patients(35 ears) who received CI in our hospital from June 2022 to July 2024 were enrolled. The Advanced Bionics Active Insertion Monitoring(AIM) system was used for real-time ECochG monitoring during surgery. Intraoperative cochlear microphonics (CM) waveform changes were recorded and analyzed in relation to postoperative RH preservation. <b>Results:</b>①ECochG recordings were successfully obtained in 34 of 35 ears (97.1%). ②According to Harris classification, there were 7 ears(20.6%) of Type A(rising), 7 ears(20.6%) of Type C(declining), 8 ears(23.5%) of Type CC(fluctuating), and 12 ears(35.3%) of Type D(no response). ③The total CM amplitude decrease was significantly moderately correlated with postoperative low-mid frequency hearing loss(<i>r</i>=0.67, <i>P</i>=0.017). The total CM amplitude decrease was significantly moderately correlated with postoperative low frequency hearing loss(<i>r</i>=0.65, <i>P</i>=0.023). ④For the mean amplitude variation, the Amax was 30.70 μV, the Amin was 8.64 μV, and the Aend was 18.27 μV. ⑤Sixteen cases completed postoperative follow-up, with an average low-mid frequency(125-1 000 Hz) residual hearing loss of 15.25 dB HL and a RH preservation rate of 87.5%. <b>Conclusion:</b>Intraoperative ECochG monitoring can effectively predict postoperative residual hearing changes, effectively guide surgical manipulation, and improve residual hearing preservation rate.

PubMedVol. 39(5)
Gene Therapy Laboratory (FR), Second People Hospital of Hunan (CN), Research International (United States) (US), Shandong Provincial Hospital (CN)
Hearing Loss and Rehabilitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.