Modification of a Two-Part Cancellous Locking Screw: A Pilot Study on Increasing Resistance to Axial Pullout Strength

<b>Background/Objectives:</b> The pullout failure of conventional locking screws (LSs, screws with a locking mechanism) may occur in patients with osteoporosis, particularly when inserted near joints or across periarticular fractures (e.g., proximal humerus). The two-part locking cancellous screw modification (TP-LCS, screws composed of two parts) in metaphyseal cancellous bone is hypothesized to increase bone purchase and holding power. This study aimed to test the hypothesized advantages of TP-LCS over LSs. <b>Methods:</b> An MTS 370 series frame with an axial/torsional load cell was used to test driving torque and axial pullout strength, following ASTM F543-07 standards. The TP-LCS group featured a newly modified screw design made from titanium alloy (Ti6Al4V), while conventional LSs (Synthes) were used for the control group. Statistical significance was assessed for selected comparisons relevant to the research objectives, including driving torque and axial pullout strength. <b>Results:</b> The driving torque test showed that TP-LCS had a significantly higher maximum insertion torque (4.9 ± 0.4 N·cm) compared to LSs (4.2 ± 0.4 N·cm) (<i>p</i> = 0.0269), although no significant difference was found in maximum removal torque (<i>p</i> = 0.1046). The axial pullout test revealed that TP-LCS had significantly higher pullout strength (223.5 ± 12.2 N) compared to LSs (203.5 ± 11.5 N) (<i>p</i> = 0.0284). Failure during the axial pullout test often involved cracking of the test block material around the screw threads, causing the screw to pull out. <b>Conclusions:</b> These results support the hypothesis that TP-LCS may offer improved axial pullout resistance compared to LSs, making it a potentially beneficial modification for LSs in osteoporotic metaphyseal regions or near joints. This study provides biomechanical insights into the advantages of the modified screw design over conventional LSs.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2025-04-23
DOI
https://doi.org/10.3390/bioengineering12050444
Primary Topic
Bone fractures and treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Modification of a Two-Part Cancellous Locking Screw: A Pilot Study on Increasing Resistance to Axial Pullout Strength

Nin‐Chieh Hsu, Hsuan‐Ti Huang, Chung‐Hwan Chen, Cheng‐Chang Lu et al.
Bioengineering
Bone fractures and treatments
article

Modification of a Two-Part Cancellous Locking Screw: A Pilot Study on Increasing Resistance to Axial Pullout Strength

Nin‐Chieh Hsu, Hsuan‐Ti Huang, Chung‐Hwan Chen, Cheng‐Chang Lu, Sung‐Yen Lin, Yin‐Chih Fu, Chia‐Hao Hsu, Pei‐Hsi Chou
article en

Abstract

<b>Background/Objectives:</b> The pullout failure of conventional locking screws (LSs, screws with a locking mechanism) may occur in patients with osteoporosis, particularly when inserted near joints or across periarticular fractures (e.g., proximal humerus). The two-part locking cancellous screw modification (TP-LCS, screws composed of two parts) in metaphyseal cancellous bone is hypothesized to increase bone purchase and holding power. This study aimed to test the hypothesized advantages of TP-LCS over LSs. <b>Methods:</b> An MTS 370 series frame with an axial/torsional load cell was used to test driving torque and axial pullout strength, following ASTM F543-07 standards. The TP-LCS group featured a newly modified screw design made from titanium alloy (Ti6Al4V), while conventional LSs (Synthes) were used for the control group. Statistical significance was assessed for selected comparisons relevant to the research objectives, including driving torque and axial pullout strength. <b>Results:</b> The driving torque test showed that TP-LCS had a significantly higher maximum insertion torque (4.9 ± 0.4 N·cm) compared to LSs (4.2 ± 0.4 N·cm) (<i>p</i> = 0.0269), although no significant difference was found in maximum removal torque (<i>p</i> = 0.1046). The axial pullout test revealed that TP-LCS had significantly higher pullout strength (223.5 ± 12.2 N) compared to LSs (203.5 ± 11.5 N) (<i>p</i> = 0.0284). Failure during the axial pullout test often involved cracking of the test block material around the screw threads, causing the screw to pull out. <b>Conclusions:</b> These results support the hypothesis that TP-LCS may offer improved axial pullout resistance compared to LSs, making it a potentially beneficial modification for LSs in osteoporotic metaphyseal regions or near joints. This study provides biomechanical insights into the advantages of the modified screw design over conventional LSs.

BioengineeringVol. 12(5)
Kaohsiung Medical University (TW), National Taiwan University (TW), Kaohsiung Medical University Chung-Ho Memorial Hospital (TW), Taipei City Hospital (TW), Kaohsiung Municipal Hsiao-Kang Hospital (TW)
Kaohsiung Municipal Ta-Tung Hospital, Kaohsiung Medical University, Kaohsiung Medical University Chung-Ho Memorial Hospital, National Science and Technology Council
Openalex Percentile: Top 88%
Bone fractures and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.