Novel method to automatize flash point detection in small volumes of liquid by computer vision using thermal images

• Novel methodology for flash point detection of 15 to 300 μL of liquid. • Closed-cup flash point apparatus combined with thermal imaging camera. • Use of computer vision for flash point detection. • Flash point detection at 95 % accuracy. A novel methodology to automate flash point detection of small solvent volumes has been successfully demonstrated and optimized. Flash point temperatures were measured using a closed-cup rapid flash point tester which was paired with a thermal imaging camera. The thermal imaging camara analyses the apparent temperature of the flame before, during and after the opening of the chamber. Two flash point standards and n-eicosane were used to validate the apparatus. Sample volume ranged between 15 and 300 μL which is of interest for the analysis of expensive or harmful liquids. This approach could eventually be extended to measuring flash points in gel polymer electrolytes, for which flammability testing is of significant interest for battery R&D. In this work, 462 flashpoints were collected to feed the machine learning algorithms. Convolutional neural network, support vector machine and random forest algorithms were used to determine the presence/absence of a flame. Flash points were predicted with an accuracy of 95 % and a precision of ±2 °C. Precision was found to be limited by the flash point detector rather than the analysis by computer vision. Other factors such as humidity (22 % to 55 %), atmospheric pressure (between 99.6 to 102.0 kPa) and volume of solvent were found to have little to no influence on flash point detection. The flash point temperatures tested in this study are limited to a range between 50 °C and 176 °C. Regression algorithms were employed to estimate the flash point temperature based on a single measurement presenting an improvement as accurate flash point detection traditionally requires several measurements.

Authors

Institutions

Publication Details

Journal
Measurement
Published
2025-04-23
DOI
https://doi.org/10.1016/j.measurement.2025.117629
Citations
2
Primary Topic
Currency Recognition and Detection
Type
article
Field-Weighted Citation Impact
2.55
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Novel method to automatize flash point detection in small volumes of liquid by computer vision using thermal images

Mickaël Dollé, Marie-Chloé Michaud Paradis, Gabrielle Y. Foran, Caroline St‐Antoine et al.
2 citations
Measurement
Currency Recognition and Detection
2.55
article

Novel method to automatize flash point detection in small volumes of liquid by computer vision using thermal images

Mickaël Dollé, Marie-Chloé Michaud Paradis, Gabrielle Y. Foran, Caroline St‐Antoine, Arnaud Prébé, Pauline Gonnel, Félix Therrien
article en
2 citations

Abstract

• Novel methodology for flash point detection of 15 to 300 μL of liquid. • Closed-cup flash point apparatus combined with thermal imaging camera. • Use of computer vision for flash point detection. • Flash point detection at 95 % accuracy. A novel methodology to automate flash point detection of small solvent volumes has been successfully demonstrated and optimized. Flash point temperatures were measured using a closed-cup rapid flash point tester which was paired with a thermal imaging camera. The thermal imaging camara analyses the apparent temperature of the flame before, during and after the opening of the chamber. Two flash point standards and n-eicosane were used to validate the apparatus. Sample volume ranged between 15 and 300 μL which is of interest for the analysis of expensive or harmful liquids. This approach could eventually be extended to measuring flash points in gel polymer electrolytes, for which flammability testing is of significant interest for battery R&D. In this work, 462 flashpoints were collected to feed the machine learning algorithms. Convolutional neural network, support vector machine and random forest algorithms were used to determine the presence/absence of a flame. Flash points were predicted with an accuracy of 95 % and a precision of ±2 °C. Precision was found to be limited by the flash point detector rather than the analysis by computer vision. Other factors such as humidity (22 % to 55 %), atmospheric pressure (between 99.6 to 102.0 kPa) and volume of solvent were found to have little to no influence on flash point detection. The flash point temperatures tested in this study are limited to a range between 50 °C and 176 °C. Regression algorithms were employed to estimate the flash point temperature based on a single measurement presenting an improvement as accurate flash point detection traditionally requires several measurements.

MeasurementVol. 253
Université du Québec à Montréal (CA), Mila - Quebec Artificial Intelligence Institute (CA), Université de Montréal (CA)
Openalex Percentile: Top 11%
Currency Recognition and Detection
2.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.