Hygrometric Performance of 113 Passive Humidifiers: ISO Standard Method Versus Psychrometry

<b>Background:</b> Passive humidifiers, also known as heat-and-moisture exchangers (HMEs) are used to warm and humidify inspired gases delivered during mechanical ventilation. Recent data demonstrate that the evaluation and selection of these devices is not optimal, leading to severe complications, such as endotracheal tube occlusions. The aim of the study was to assess the humidification performances of a large number of commercially available devices, with the psychrometric method, and to compare the results with manufacturer's data. <b>Methods:</b> We assessed 113 devices using a bench test that simulated physiological ventilation conditions. Seventy-one devices were described by the manufacturers as HME without or with filtration properties (HMEF) and 42 as antimicrobial filters. Among the 71 HME/HMEF, 60% were foam-based, 32% were paper-based, 6% were fiber-based, and 1 had carbon media (HMEF with inhaled sedation function). Three hygrometric measurements using the psychrometric method for each device were performed after reaching a steady state and compared with manufacturers' data (based on the International Organization for Standardization [ISO] method). <b>Results:</b> Among the 71 HME/HMEF tested, only 24 (34%) delivered absolute humidity above 28 mg H<sub>2</sub>O/L. Nineteen percent of the foam-based and 57% of the paper-based HME/HMEF delivered desired levels of humidity. Humidification data provided by manufacturers were available for 59 HME/HMEFs. In all but one device, humidity output was overestimated using the ISO method. The median (interquartile) difference between our measurements and the manufacturers' data were 6.0 (3.8-8.7) mg H<sub>2</sub>O/L for devices described as HMEs (<i>P</i> < .001). Poor performing devices were detected only with the psychrometric method. <b>Conclusions:</b> Several HME/HMEFs performed poorly and should not be used for prolonged mechanical ventilation. The values determined by independent assessments were lower than values reported by manufacturers. Evaluation of a passive humidifier using the current ISO method does not guarantee provision of adequate humidification.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2025-04-21
DOI
https://doi.org/10.1089/respcare.12660
Citations
4
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
6.07
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hygrometric Performance of 113 Passive Humidifiers: ISO Standard Method Versus Psychrometry

Richard D. Branson, Jean-Claude Lefebvre, Laura-Ann Cyr, François Lellouche et al.
4 citations
Respiratory Care
Respiratory Support and Mechanisms
6.07
article

Hygrometric Performance of 113 Passive Humidifiers: ISO Standard Method Versus Psychrometry

Richard D. Branson, Jean-Claude Lefebvre, Laura-Ann Cyr, François Lellouche, Pierre-Alexandre Bouchard
article en
4 citations

Abstract

<b>Background:</b> Passive humidifiers, also known as heat-and-moisture exchangers (HMEs) are used to warm and humidify inspired gases delivered during mechanical ventilation. Recent data demonstrate that the evaluation and selection of these devices is not optimal, leading to severe complications, such as endotracheal tube occlusions. The aim of the study was to assess the humidification performances of a large number of commercially available devices, with the psychrometric method, and to compare the results with manufacturer's data. <b>Methods:</b> We assessed 113 devices using a bench test that simulated physiological ventilation conditions. Seventy-one devices were described by the manufacturers as HME without or with filtration properties (HMEF) and 42 as antimicrobial filters. Among the 71 HME/HMEF, 60% were foam-based, 32% were paper-based, 6% were fiber-based, and 1 had carbon media (HMEF with inhaled sedation function). Three hygrometric measurements using the psychrometric method for each device were performed after reaching a steady state and compared with manufacturers' data (based on the International Organization for Standardization [ISO] method). <b>Results:</b> Among the 71 HME/HMEF tested, only 24 (34%) delivered absolute humidity above 28 mg H<sub>2</sub>O/L. Nineteen percent of the foam-based and 57% of the paper-based HME/HMEF delivered desired levels of humidity. Humidification data provided by manufacturers were available for 59 HME/HMEFs. In all but one device, humidity output was overestimated using the ISO method. The median (interquartile) difference between our measurements and the manufacturers' data were 6.0 (3.8-8.7) mg H<sub>2</sub>O/L for devices described as HMEs (<i>P</i> < .001). Poor performing devices were detected only with the psychrometric method. <b>Conclusions:</b> Several HME/HMEFs performed poorly and should not be used for prolonged mechanical ventilation. The values determined by independent assessments were lower than values reported by manufacturers. Evaluation of a passive humidifier using the current ISO method does not guarantee provision of adequate humidification.

Respiratory CareVol. 70(11)
Institut universitaire de cardiologie et de pneumologie de Québec (CA), Université Laval (CA), University of Cincinnati (US), University of Cincinnati Medical Center (US)
Climate action
Openalex Percentile: Top 4%
Respiratory Support and Mechanisms
6.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.