The Influence of Pressurized Metered Dose Inhaler Actuator Orifice Diameter on Regional Extrathoracic Deposition of Inhaled Epinephrine

<b><i>Background:</i></b> Extrathoracic deposition is a large source of <i>in vivo</i> variability in dosing for pressurized metered dose inhaler (pMDI) aerosols. A majority of previous studies have focused on only total extrathoracic deposition for pMDIs. The present work evaluates regional deposition within the extrathoracic region to better understand the impact of actuator orifice diameter and inhalation flow rate on extrathoracic deposition of a suspension pMDI formulation of epinephrine. <b><i>Methods:</i></b> Regional deposition of a commercially available HFA (hydrofluoroalkane) suspension pMDI formulation of epinephrine was evaluated using plastic and metal versions of the newly developed sectioned Alberta Idealized Throat (s-AIT), divided into analogs of the oral cavity, the pharynx/larynx, and the upper trachea. Influences of actuator orifice diameter and inhaler insertion angle on regional extrathoracic deposition were evaluated in the plastic s-AIT at a 30 L/min flow rate, followed by additional testing in the metal s-AIT to evaluate effects across a range of flow rates (from 10 to 100 L/min). <b><i>Results:</i></b> Actuator orifice was found to strongly influence regional extrathoracic deposition of a commercially available epinephrine HFA suspension pMDI aerosol, with smaller actuator orifices yielding reduced oral cavity deposition and increased distal-filter (<i>in vitro</i> lung) deposition in both the plastic and metal s-AIT. Inhalation flow rate was found to strongly influence deposition in the metal s-AIT, with higher flow rates associated with reduced oral cavity deposition, increased pharynx/larynx deposition, and increased upper trachea deposition. Smaller orifices showed less variability in results as a function of inhaler insertion angle. <b><i>Conclusions:</i></b> Actuator orifice diameter (spanning 0.22-0.42 mm) can strongly influence regional deposition of an HFA epinephrine suspension pMDI aerosol within the extrathoracic region. Smaller actuator orifices may provide reduced oral cavity deposition and increased delivery to the lungs. Smaller actuator orifices may also reduce variability in extrathoracic deposition that is associated with patient use aspects such as inhaler insertion angle.

Authors

Institutions

Publication Details

Journal
Journal of Aerosol Medicine and Pulmonary Drug Delivery
Published
2025-04-21
DOI
https://doi.org/10.1089/jamp.2024.0039
Citations
1
Primary Topic
Inhalation and Respiratory Drug Delivery
Type
article
Field-Weighted Citation Impact
1.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Influence of Pressurized Metered Dose Inhaler Actuator Orifice Diameter on Regional Extrathoracic Deposition of Inhaled Epinephrine

K.W. STAPLETON, Conor A. Ruzycki, George H. Luciuk, Andrew R. Martin et al.
1 citations
Journal of Aerosol Medicine and Pulmonary Drug Delivery
Inhalation and Respiratory Drug Delivery
1.52
article

The Influence of Pressurized Metered Dose Inhaler Actuator Orifice Diameter on Regional Extrathoracic Deposition of Inhaled Epinephrine

K.W. STAPLETON, Conor A. Ruzycki, George H. Luciuk, Andrew R. Martin, Scott Tavernini, Warren H. Finlay
article en
1 citations

Abstract

<b><i>Background:</i></b> Extrathoracic deposition is a large source of <i>in vivo</i> variability in dosing for pressurized metered dose inhaler (pMDI) aerosols. A majority of previous studies have focused on only total extrathoracic deposition for pMDIs. The present work evaluates regional deposition within the extrathoracic region to better understand the impact of actuator orifice diameter and inhalation flow rate on extrathoracic deposition of a suspension pMDI formulation of epinephrine. <b><i>Methods:</i></b> Regional deposition of a commercially available HFA (hydrofluoroalkane) suspension pMDI formulation of epinephrine was evaluated using plastic and metal versions of the newly developed sectioned Alberta Idealized Throat (s-AIT), divided into analogs of the oral cavity, the pharynx/larynx, and the upper trachea. Influences of actuator orifice diameter and inhaler insertion angle on regional extrathoracic deposition were evaluated in the plastic s-AIT at a 30 L/min flow rate, followed by additional testing in the metal s-AIT to evaluate effects across a range of flow rates (from 10 to 100 L/min). <b><i>Results:</i></b> Actuator orifice was found to strongly influence regional extrathoracic deposition of a commercially available epinephrine HFA suspension pMDI aerosol, with smaller actuator orifices yielding reduced oral cavity deposition and increased distal-filter (<i>in vitro</i> lung) deposition in both the plastic and metal s-AIT. Inhalation flow rate was found to strongly influence deposition in the metal s-AIT, with higher flow rates associated with reduced oral cavity deposition, increased pharynx/larynx deposition, and increased upper trachea deposition. Smaller orifices showed less variability in results as a function of inhaler insertion angle. <b><i>Conclusions:</i></b> Actuator orifice diameter (spanning 0.22-0.42 mm) can strongly influence regional deposition of an HFA epinephrine suspension pMDI aerosol within the extrathoracic region. Smaller actuator orifices may provide reduced oral cavity deposition and increased delivery to the lungs. Smaller actuator orifices may also reduce variability in extrathoracic deposition that is associated with patient use aspects such as inhaler insertion angle.

Journal of Aerosol Medicine and Pulmonary Drug DeliveryVol. 38(3)
University of Alberta (CA), Lovelace Medical Center (US), Supratek Pharma (Canada) (CA)
Clean water and sanitation
Openalex Percentile: Top 20%
Inhalation and Respiratory Drug Delivery
1.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.