CO 2 gas sensors are adopted in various manufacturing industries, medical sectors and also for indoor air quality measurement. Their functions range from concentration monitoring for product quality to detecting hazardous levels for safety. Having a robust CO 2 gas sensor is important, but these sensors can be affected by temperature and humidity in the environment. Here we demonstrate an NDIR CO 2 gas sensor based on 20 % ScAlN-based pyroelectric detector chip, tested under different environmental conditions. Despite absolute signal output drop when surrounding temperatures increase, measured results show that CO 2 gas response remains the same. Moreover, relative humidity (RH) does not seem to have much effect on the response of this NDIR CO 2 gas sensor. Response times stay at ∼1–2 s despite changes in environmental conditions. This sensor is also tested continuously for 100 hours to observe its signal variation over long hours. We note signal variation of ∼± 0.44 % during the first 40 hours with subsequent variation increase to ∼ ± 1.7 % at the 50th hour. Robustness of CO 2 gas sensing with 20 % ScAlN-based pyroelectric detector chip is also observed by measuring the gas responses of 5 detector chips obtained from 5 different locations within an 8-inch wafer, measuring CO 2 gas sensing performance uniformity ∼0.48 % within an 8-inch wafer. The results obtained will allow us to have further insights and understanding on the potential and practicality of NDIR gas sensors based on 20 % ScAlN-based pyroelectric detectors when subjected to different environmental conditions. • CO2 gas sensing based on 20 % ScAlN-based pyroelectric detector. • CO2 gas sensor demonstrates robustness to relative humidity and temperature change. • Response and recovery times stay at 1–2 seconds. • Sensing signal variation within ∼ ± 0.44 % during the first 40 hours. • Detectors’ performance uniformity across 8-inch wafer ∼0.48 %.

Authors

Institutions

Publication Details

Journal
Sensors and Actuators B Chemical
Published
2025-04-21
DOI
https://doi.org/10.1016/j.snb.2025.137843
Citations
3
Primary Topic
Gas Sensing Nanomaterials and Sensors
Type
article
Field-Weighted Citation Impact
2.19

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Environmental influences on NDIR CO2 gas sensor using 20 % ScAlN-based pyroelectric detector chip

Zhonghua Gu, Chong Pei Ho, Doris K. T. Ng, Md Hazwani Khairy Md Husni et al.
3 citations
Sensors and Actuators B Chemical
Gas Sensing Nanomaterials and Sensors
2.19
article

Environmental influences on NDIR CO2 gas sensor using 20 % ScAlN-based pyroelectric detector chip

Zhonghua Gu, Chong Pei Ho, Doris K. T. Ng, Md Hazwani Khairy Md Husni, Norhanani Jaafar, Jia Sheng Goh, Kristel Pei Xuan Wee, Huanhuan Wang, Weiguo Chen, Qingxin Zhang, Linfang Xu, William Provenaz
article en
3 citations

Abstract

CO 2 gas sensors are adopted in various manufacturing industries, medical sectors and also for indoor air quality measurement. Their functions range from concentration monitoring for product quality to detecting hazardous levels for safety. Having a robust CO 2 gas sensor is important, but these sensors can be affected by temperature and humidity in the environment. Here we demonstrate an NDIR CO 2 gas sensor based on 20 % ScAlN-based pyroelectric detector chip, tested under different environmental conditions. Despite absolute signal output drop when surrounding temperatures increase, measured results show that CO 2 gas response remains the same. Moreover, relative humidity (RH) does not seem to have much effect on the response of this NDIR CO 2 gas sensor. Response times stay at ∼1–2 s despite changes in environmental conditions. This sensor is also tested continuously for 100 hours to observe its signal variation over long hours. We note signal variation of ∼± 0.44 % during the first 40 hours with subsequent variation increase to ∼ ± 1.7 % at the 50th hour. Robustness of CO 2 gas sensing with 20 % ScAlN-based pyroelectric detector chip is also observed by measuring the gas responses of 5 detector chips obtained from 5 different locations within an 8-inch wafer, measuring CO 2 gas sensing performance uniformity ∼0.48 % within an 8-inch wafer. The results obtained will allow us to have further insights and understanding on the potential and practicality of NDIR gas sensors based on 20 % ScAlN-based pyroelectric detectors when subjected to different environmental conditions. • CO2 gas sensing based on 20 % ScAlN-based pyroelectric detector. • CO2 gas sensor demonstrates robustness to relative humidity and temperature change. • Response and recovery times stay at 1–2 seconds. • Sensing signal variation within ∼ ± 0.44 % during the first 40 hours. • Detectors’ performance uniformity across 8-inch wafer ∼0.48 %.

Sensors and Actuators B ChemicalVol. 439
Agency for Science, Technology and Research (SG), Institute of Materials Research and Engineering (SG)
Agency for Science, Technology and Research, National Research Foundation Singapore
Openalex Percentile: Top 13%
Gas Sensing Nanomaterials and Sensors
2.19
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.