Abstract 1335: Comparison of X-ray and gamma-ray irradiators for in vitro and in vivo biological effectiveness

Abstract Objective: The Cesium Irradiator Replacement Program (CIRP) was established in 2014 to replace radioisotope-based irradiators with non-radioactive alternatives, such as X-rays. As most dosimetry is based on 137Cs irradiators, transitioning to X-ray systems has presented challenges in radiobiological research. A key obstacle stems from the inaccurate assumption of dose as a directly transferable constant across different irradiation types. To mitigate this issue, the University of Wisconsin-Madison, as a CIRP participant, has undertaken a study to establish consistency in dose measurement and biological response, both in vitro and in vivo, with a specific emphasis on the relative biological effectiveness of X-rays compared to cesium. Methods: Radiation dose was calibrated using experimental phantoms and thermoluminescent dosimeters. To determine dose response in vitro, HeLa cells were treated with 0 to 8 Gy of radiation in 2 Gy increments delivered from a 137Cs, 60Co, or 195 kVp X-ray source. Colony formation and comet assays were used to evaluate clonogenic survival and DNA damage, respectively. For in vivo studies, C57BL/6 mice received a single 8 Gy whole-body radiation dose from a 137Cs, 60Co, 220 kVp, or 300 kVp X-ray source. Lethal Dose50/30 (LD50/30) was used as the primary end point. Histopathological analysis of the spleen and bone marrow was performed to identify primary radiation-induced organ damage. Quantitative assessment of tissue injury was conducted using Halo image analysis. Hematology and blood chemistry were measured to monitor radiation-induced toxicity and subsequent recovery. Results: The 195 kVp x-ray source produced the highest number of DNA double-strand breaks, followed by 137Cs and 60Co. This was most evident as the total radiation dose increased and translated to a greater impairment of clonogenic survival, particularly evident at 4, 6, and 8 Gy for 195 kVp X-rays. While 137Cs and 60Co had similar effects at lower doses, 137Cs induced a more pronounced reduction in colony formation at 6 and 8 Gy. In animal studies, 60Co exhibited the highest LD50/30 (8.5 Gy), followed by 137Cs (8.1 Gy), 300 kVp (7.3 Gy), and 220 kVp (6.2 Gy), respectively. Conclusion: The differences were found in the biological effects between radioisotope-based and X-ray irradiators when delivered the identical physical dose. In animal models, radiation from a 60C source demonstrated a weaker biological effect than 137Cs, while 220 kVp X-rays showed the strongest biological effect. Given the differences in the energy spectra of the radiation sources, we will further investigate how dose-energy distribution influences biological effects in radiation-sensitive organs, particularly superficial tissues and hematopoietic systems. Citation Format: Caroline Mae Mork, Shrey Ramesh, Keith Kunugi, Kwangok P. Nickel, Larry Dewerd, Nina Zitzer, Randall Kimple. Comparison of X-ray and gamma-ray irradiators for in vitro and in vivo biological effectiveness [abstract]. In: Proceedings of the American Association for Cancer Research Annual Meeting 2025; Part 1 (Regular Abstracts); 2025 Apr 25-30; Chicago, IL. Philadelphia (PA): AACR; Cancer Res 2025;85(8_Suppl_1):Abstract nr 1335.

Authors

Institutions

Publication Details

Journal
Cancer Research
Published
2025-04-21
DOI
https://doi.org/10.1158/1538-7445.am2025-1335
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 1335: Comparison of X-ray and gamma-ray irradiators for in vitro and in vivo biological effectiveness

Cato Mørk, Shrey Ramesh, Nina C. Zitzer, Larry A. DeWerd et al.
Cancer Research
Ultrasound and Hyperthermia Applications
article

Abstract 1335: Comparison of X-ray and gamma-ray irradiators for in vitro and in vivo biological effectiveness

Cato Mørk, Shrey Ramesh, Nina C. Zitzer, Larry A. DeWerd, Randall J. Kimple, Keith A. Kunugi, Kwangok P. Nickel
article en

Abstract

Abstract Objective: The Cesium Irradiator Replacement Program (CIRP) was established in 2014 to replace radioisotope-based irradiators with non-radioactive alternatives, such as X-rays. As most dosimetry is based on 137Cs irradiators, transitioning to X-ray systems has presented challenges in radiobiological research. A key obstacle stems from the inaccurate assumption of dose as a directly transferable constant across different irradiation types. To mitigate this issue, the University of Wisconsin-Madison, as a CIRP participant, has undertaken a study to establish consistency in dose measurement and biological response, both in vitro and in vivo, with a specific emphasis on the relative biological effectiveness of X-rays compared to cesium. Methods: Radiation dose was calibrated using experimental phantoms and thermoluminescent dosimeters. To determine dose response in vitro, HeLa cells were treated with 0 to 8 Gy of radiation in 2 Gy increments delivered from a 137Cs, 60Co, or 195 kVp X-ray source. Colony formation and comet assays were used to evaluate clonogenic survival and DNA damage, respectively. For in vivo studies, C57BL/6 mice received a single 8 Gy whole-body radiation dose from a 137Cs, 60Co, 220 kVp, or 300 kVp X-ray source. Lethal Dose50/30 (LD50/30) was used as the primary end point. Histopathological analysis of the spleen and bone marrow was performed to identify primary radiation-induced organ damage. Quantitative assessment of tissue injury was conducted using Halo image analysis. Hematology and blood chemistry were measured to monitor radiation-induced toxicity and subsequent recovery. Results: The 195 kVp x-ray source produced the highest number of DNA double-strand breaks, followed by 137Cs and 60Co. This was most evident as the total radiation dose increased and translated to a greater impairment of clonogenic survival, particularly evident at 4, 6, and 8 Gy for 195 kVp X-rays. While 137Cs and 60Co had similar effects at lower doses, 137Cs induced a more pronounced reduction in colony formation at 6 and 8 Gy. In animal studies, 60Co exhibited the highest LD50/30 (8.5 Gy), followed by 137Cs (8.1 Gy), 300 kVp (7.3 Gy), and 220 kVp (6.2 Gy), respectively. Conclusion: The differences were found in the biological effects between radioisotope-based and X-ray irradiators when delivered the identical physical dose. In animal models, radiation from a 60C source demonstrated a weaker biological effect than 137Cs, while 220 kVp X-rays showed the strongest biological effect. Given the differences in the energy spectra of the radiation sources, we will further investigate how dose-energy distribution influences biological effects in radiation-sensitive organs, particularly superficial tissues and hematopoietic systems. Citation Format: Caroline Mae Mork, Shrey Ramesh, Keith Kunugi, Kwangok P. Nickel, Larry Dewerd, Nina Zitzer, Randall Kimple. Comparison of X-ray and gamma-ray irradiators for in vitro and in vivo biological effectiveness [abstract]. In: Proceedings of the American Association for Cancer Research Annual Meeting 2025; Part 1 (Regular Abstracts); 2025 Apr 25-30; Chicago, IL. Philadelphia (PA): AACR; Cancer Res 2025;85(8_Suppl_1):Abstract nr 1335.

Cancer ResearchVol. 85(8_Supplement_1)
University of Wisconsin–Madison (US), University of Wisconsin Carbone Cancer Center (US)
Openalex Percentile: Top 93%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.