Segmentation of hippocampal subregion via 5-Tesla magnetic resonance imaging: a comparative study
The 5-T magnetic resonance system offers higher SNR and CNR values for hippocampal imaging as compared to 3-T imaging. A slice thickness of 0.7-1 mm is recommended for 2D T2-weighted imaging. However, in the selection of automatic segmentation tools for hippocampal subregions, caution is advised if the slice thickness is less than 2 mm.
Authors
- Dong Liang (ORCID: https://orcid.org/0000-0001-6257-0875)
- Yijun Zhou (ORCID: https://orcid.org/0000-0001-9593-4655)
- Ye Li (ORCID: https://orcid.org/0000-0002-4585-956X)
- Xin Liu (ORCID: https://orcid.org/0000-0001-8045-8254)
- Yining Wang (ORCID: https://orcid.org/0000-0001-6441-2002)
- Shuo Chen
- Jun Ni (ORCID: https://orcid.org/0000-0002-8135-6951)
- Lixian Zou
- Hairong Zheng
Institutions
- Chinese Academy of Medical Sciences & Peking Union Medical College (CN)
- Peking Union Medical College Hospital (CN)
- University Town of Shenzhen (CN)
- United Imaging Healthcare (China) (CN)
- Shenzhen Institutes of Advanced Technology (CN)
- Shenzhen Technology University (CN)
Publication Details
- Journal
- Quantitative Imaging in Medicine and Surgery
- Published
- 2025-04-21
- DOI
- https://doi.org/10.21037/qims-24-2169
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00