Development of a Smartphone-Linked Immunosensing System for Oxytocin Determination
We report an optical immunosensing system for oxytocin (OXT) based on image analysis of color reactions in an enzyme-linked immunosorbent assay (ELISA). We employed a miniaturized optical immunosensing unit that was functionally connected to an LED and a smartphone camera. Our system measures OXT levels using a metric called the RGBscore, which is derived from the red, green, and blue (RGB) information in the captured images. By calculating the RGBscore regressively using the brute-force method, this approach can be applied to smartphones with various CMOS image sensors and firmware. The lower detection limit was determined to be 5.26 pg/mL, and the measurement results showed a higher correlation (r = 0.972) with those obtained from conventional ELISA. These results suggest the potential for its application in a simplified health management system for individuals.
Authors
- Hiroyuki Kudo (ORCID: https://orcid.org/0000-0001-7256-6554)
- Miku Sarubo
- Yuka Numazaki
- Yusuke Suzuki
Institutions
- Meiji University (JP)
Publication Details
- Journal
- Biosensors
- Published
- 2025-04-18
- DOI
- https://doi.org/10.3390/bios15040261
- Citations
- 1
- Primary Topic
- Neuroendocrine regulation and behavior
- Type
- article
- Field-Weighted Citation Impact
- 2.44
Funders
- Japan Society for the Promotion of Science