Comparing the Efficiency of Valved Trocar Cannulas for Pars Plana Vitrectomy
Despite significant differences, all tested valved cannulas satisfy safety criteria by keeping a surgically negligible difference between the set and actual intraocular pressure. The minimal leakage measure may act as a safety pressure damper under critical conditions.