The validation of urea determination in saliva using optical urea biosensor with p-dimethylaminobenzaldehyde (DMAB) reagent

The validation of urea determination in saliva using the biosensor method and the p-dimethylaminobenzaldehyde (DMAB) method has been successfully conducted. Anthocyanin compounds from Chatarantus roseus flowers were extracted using the maceration method with methanol as the solvent, yielding a total of 22.60% with a total concentration of 6.01 mg/L. The obtained extract was subjected to qualitative anthocyanin testing, showing a positive result was indicated by the formation of a reddish color and quantitative testing with the formation of a faded yellow color. The maximum wavelength (λ max) of anthocyanin was 664 nm with an absorbance of 0.674. The saliva samples were obtained from three different ages i.e: children, teenagers, and adults. During the saliva collection procedure, participants were instructed to abstain from consuming any food or beverages, with the exception of water, for a minimum of 1 hour prior to sample collection. All samples were stored in a freezer at 4-8°C until needed for analysis. Linearity test results were assessed using a calibration curve, yielding a coefficient of determination (R²) of 0.978 with a sensitivity of 0.033 for the biosensor and an R² of 0.975 with a sensitivity of 0.685 for DMAB. The limit of detection (LOD) values were determined as 7.203 × 10⁻¹¹ M for the biosensor and 6.984 × 10⁻⁶ M for DMAB, while the limit of quantification (LOQ) values were 2.182 × 10⁻¹⁰ M and 2.116 × 10⁻⁵ M, respectively. Statistical analysis using a t-test showed tcalculated = 1.314, compared to ttable = 4.302, indicating that tcalculated ttable. This result suggests no statistically significant difference between the biosensor and DMAB methods, confirming that the biosensor method is comparable in performance to the DMAB method. Additionally, urea concentration measurements showed that adult samples exhibited the highest urea levels among the tested samples, which may indicate an association with potential health risks, including dental caries, kidney failure, and liver damage.

Authors

Publication Details

Journal
Jurnal Natural
Published
2025-03-05
DOI
https://doi.org/10.24815/jn.v25i1.41251
Primary Topic
Salivary Gland Disorders and Functions
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The validation of urea determination in saliva using optical urea biosensor with p-dimethylaminobenzaldehyde (DMAB) reagent

Binawati Ginting, Eka Safitri, Salsabilla Latansa Nazaruddin, Khairi Suhud et al.
Jurnal Natural
Salivary Gland Disorders and Functions
article

The validation of urea determination in saliva using optical urea biosensor with p-dimethylaminobenzaldehyde (DMAB) reagent

Binawati Ginting, Eka Safitri, Salsabilla Latansa Nazaruddin, Khairi Suhud, Muhammad Afifi, Nurhaida Nurhaida, Sitti Saleha, Teuku Aldiansyah, Nurhayati Nurhayati, Nazaruddin Nazaruddin
article en

Abstract

The validation of urea determination in saliva using the biosensor method and the p-dimethylaminobenzaldehyde (DMAB) method has been successfully conducted. Anthocyanin compounds from Chatarantus roseus flowers were extracted using the maceration method with methanol as the solvent, yielding a total of 22.60% with a total concentration of 6.01 mg/L. The obtained extract was subjected to qualitative anthocyanin testing, showing a positive result was indicated by the formation of a reddish color and quantitative testing with the formation of a faded yellow color. The maximum wavelength (λ max) of anthocyanin was 664 nm with an absorbance of 0.674. The saliva samples were obtained from three different ages i.e: children, teenagers, and adults. During the saliva collection procedure, participants were instructed to abstain from consuming any food or beverages, with the exception of water, for a minimum of 1 hour prior to sample collection. All samples were stored in a freezer at 4-8°C until needed for analysis. Linearity test results were assessed using a calibration curve, yielding a coefficient of determination (R²) of 0.978 with a sensitivity of 0.033 for the biosensor and an R² of 0.975 with a sensitivity of 0.685 for DMAB. The limit of detection (LOD) values were determined as 7.203 × 10⁻¹¹ M for the biosensor and 6.984 × 10⁻⁶ M for DMAB, while the limit of quantification (LOQ) values were 2.182 × 10⁻¹⁰ M and 2.116 × 10⁻⁵ M, respectively. Statistical analysis using a t-test showed tcalculated = 1.314, compared to ttable = 4.302, indicating that tcalculated ttable. This result suggests no statistically significant difference between the biosensor and DMAB methods, confirming that the biosensor method is comparable in performance to the DMAB method. Additionally, urea concentration measurements showed that adult samples exhibited the highest urea levels among the tested samples, which may indicate an association with potential health risks, including dental caries, kidney failure, and liver damage.

Jurnal NaturalVol. 25(1)
Openalex Percentile: Top 90%
Salivary Gland Disorders and Functions
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.