Dynamic response of soft tissue can be disregarded during femoral stem impaction
Common metal-on-metal mallet blows in surgery are very short and mostly excite high frequencies that are clearly above the natural frequency of the femur-tissue system. Those overcritical force impulses make the stem slide into the femur before the bone can even start moving. Hence, the individual mechanical behavior of the femur-tissue system can be disregarded provided that the force is applied with very short blows. This needs to be considered for any attempt to replace the mallet in the operation theater (e.g. automated surgical impaction tools) or to modify the mallet (e.g. alternative tip material). Furthermore, it may provide guidance on the fixation of femurs in in vitro testing to mimic surgical reality.