Comparison of Effective Dose From Intraoperative 3D Imaging Using Mobile and Fixed C-Arms Versus Computed Tomography When Evaluating the Lumbar Spine

Study designExperimental Study.ObjectivesIntraoperative three-dimensional (3D) imaging has been shown to increase patient safety by assuring proper implant placement and often makes postoperative CT imaging obsolete. Nonetheless, there is no data about effective radiation doses in comparison to CT imaging.MethodsEffective doses of 3D and CT scans were measured with an anthropomorphic phantom (CIRS Inc, ATOM® Dosimetry Phantoms, Norfolk, VA, USA). Clinical presets of the C-arms were used and measured separately (mobile: Cios Spin; fixed: ARTIS pheno; both Siemens Healthcare GmbH, Erlangen, Germany). For CT imaging, a protocol for high-contrast imaging was used (SOMATOM X.cite, Siemens Healthcare GmbH, Erlangen, Germany).ResultsDepending on presets, effective doses with the mobile C-arm ranged from .51 ± .02 mSv (dose-sparing preset) to 3.14 ± .01 mSv (high-quality preset, <i>P</i> < .001). With the fixed C-arm, the effective dose ranged from .52 ± .0 mSv (dose-sparing preset) to 1.6 ± .02 mSv (standard-quality preset, <i>P</i> < .001). With CT imaging, the effective dose was 2.0 mSv.ConclusionsWhen evaluating the lumbar spine, the effective dose from intraoperative 3D imaging using mobile and fixed C-arms is comparable to that from CT scans.Proper choice of intraoperative protocol is essential, as there can be up to a 6-fold variation in dose between protocols. Intraoperative 3D imaging with (mobile or fixed) C-arms has been proven to avoid revision surgery. In this study, effective doses were shown to be comparable to those of CT scans. To avoid unnecessary exposure, indications need to be clear, and the choice of acquiring settings must respect the ALARA principle ("as low as reasonably achievable").

Authors

Institutions

Publication Details

Journal
Global Spine Journal
Published
2025-04-17
DOI
https://doi.org/10.1177/21925682251334986
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Effective Dose From Intraoperative 3D Imaging Using Mobile and Fixed C-Arms Versus Computed Tomography When Evaluating the Lumbar Spine

Marcus Brehm, E. Stepina, Mario Perl, Holger Stadthalter et al.
Global Spine Journal
Radiation Dose and Imaging
article

Comparison of Effective Dose From Intraoperative 3D Imaging Using Mobile and Fixed C-Arms Versus Computed Tomography When Evaluating the Lumbar Spine

Marcus Brehm, E. Stepina, Mario Perl, Holger Stadthalter, A. Grafenberg, U. Liebegut
article en

Abstract

Study designExperimental Study.ObjectivesIntraoperative three-dimensional (3D) imaging has been shown to increase patient safety by assuring proper implant placement and often makes postoperative CT imaging obsolete. Nonetheless, there is no data about effective radiation doses in comparison to CT imaging.MethodsEffective doses of 3D and CT scans were measured with an anthropomorphic phantom (CIRS Inc, ATOM® Dosimetry Phantoms, Norfolk, VA, USA). Clinical presets of the C-arms were used and measured separately (mobile: Cios Spin; fixed: ARTIS pheno; both Siemens Healthcare GmbH, Erlangen, Germany). For CT imaging, a protocol for high-contrast imaging was used (SOMATOM X.cite, Siemens Healthcare GmbH, Erlangen, Germany).ResultsDepending on presets, effective doses with the mobile C-arm ranged from .51 ± .02 mSv (dose-sparing preset) to 3.14 ± .01 mSv (high-quality preset, <i>P</i> < .001). With the fixed C-arm, the effective dose ranged from .52 ± .0 mSv (dose-sparing preset) to 1.6 ± .02 mSv (standard-quality preset, <i>P</i> < .001). With CT imaging, the effective dose was 2.0 mSv.ConclusionsWhen evaluating the lumbar spine, the effective dose from intraoperative 3D imaging using mobile and fixed C-arms is comparable to that from CT scans.Proper choice of intraoperative protocol is essential, as there can be up to a 6-fold variation in dose between protocols. Intraoperative 3D imaging with (mobile or fixed) C-arms has been proven to avoid revision surgery. In this study, effective doses were shown to be comparable to those of CT scans. To avoid unnecessary exposure, indications need to be clear, and the choice of acquiring settings must respect the ALARA principle ("as low as reasonably achievable").

Global Spine JournalVol. 15(8)
Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Universitätsklinikum Erlangen (DE), Siemens Healthcare (United States) (US), Siemens Healthineers (Germany) (DE)
Siemens Healthineers
Openalex Percentile: Top 88%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.