Cathodoluminescence in a Scanning Electron Microscope Operated in Transmission Mode

Abstract This study demonstrates the potential of correlative analysis cathodoluminescence (CL) carried out in a scanning electron microscope (SEM) operated in transmission mode in combination with a segmented detector. As demonstrated for the analysis of an (Al,Ga)N/AlN layer system, the use of 30 keV primary electrons permits the correlative recording of scanning transmission electron microscopy (STEM) bright field (BF), dark field (DF), high-angle annular dark field (HAADF), along with the CL signal. Despite the limitation to 30 keV, the transmitted electron signals provide an enhanced signal-to-noise ratio and improved chemical sensitivity as compared to any in-lens or chamber detector. Quantitative evaluation of the BF signal facilitates the estimation of sample thickness and identification of extended defects. The HAADF signal is significantly influenced by dynamical scattering effects, rendering material contrast highly thickness-dependent. Consequently, quantitative compositional analysis requires precise knowledge of the sample thickness, or alternatively, correlative analysis with spectrally resolved CL.

Authors

Institutions

Publication Details

Journal
Journal of Electronic Materials
Published
2025-04-18
DOI
https://doi.org/10.1007/s11664-025-11904-9
Primary Topic
Electron and X-Ray Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Cathodoluminescence in a Scanning Electron Microscope Operated in Transmission Mode

A. Lachowski, Robert Kernke, Tobias Schulz, M. Albrecht
Journal of Electronic Materials
Electron and X-Ray Spectroscopy Techniques
article

Cathodoluminescence in a Scanning Electron Microscope Operated in Transmission Mode

A. Lachowski, Robert Kernke, Tobias Schulz, M. Albrecht
article en

Abstract

Abstract This study demonstrates the potential of correlative analysis cathodoluminescence (CL) carried out in a scanning electron microscope (SEM) operated in transmission mode in combination with a segmented detector. As demonstrated for the analysis of an (Al,Ga)N/AlN layer system, the use of 30 keV primary electrons permits the correlative recording of scanning transmission electron microscopy (STEM) bright field (BF), dark field (DF), high-angle annular dark field (HAADF), along with the CL signal. Despite the limitation to 30 keV, the transmitted electron signals provide an enhanced signal-to-noise ratio and improved chemical sensitivity as compared to any in-lens or chamber detector. Quantitative evaluation of the BF signal facilitates the estimation of sample thickness and identification of extended defects. The HAADF signal is significantly influenced by dynamical scattering effects, rendering material contrast highly thickness-dependent. Consequently, quantitative compositional analysis requires precise knowledge of the sample thickness, or alternatively, correlative analysis with spectrally resolved CL.

Journal of Electronic MaterialsVol. 54(7)
Leibniz Institute for Crystal Growth (DE)
Openalex Percentile: Top 92%
Electron and X-Ray Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.