Accuracy of infusion flow rates and bolus doses for portable infusion pump

This study aimed to investigate the flow rate accuracy and bolus doses of four portable pumps with different mechanisms. We evaluated two electronic (COOPDECH Amy PCA [Amy] and CADD-Solis) and two elastomeric balloon-type pumps (COOPDECH Balloonjector and Rakurakufuser) under the following conditions: placement with an epidural catheter, operation on a shaking table, warming to 32 °C, and at room temperature (control, 25 °C). Amy maintained consistent flow rates across all conditions. The CADD-Solis also exhibited consistent flow rates, with minor yet statistically significant changes upon epidural catheter placement (- 5.0%, P = 0.005) and motion conditions (4.2%, P = 0.015). The Balloonjector and Rakurakufuser exhibited flow rate variations over time, and temperature increases significantly increased flow rates by 24.3% (P < 0.001) and 20.3% (P < 0.001), respectively. Bolus volume accuracy for the Amy, Balloonjector, and Rakurakufuser was not significantly affected under different conditions. The CADD-Solis showed a slight decrease in bolus volume with an epidural catheter (mean difference, 0.3 mL; P = 0.003). The electronic pumps maintained consistent flow rates across various conditions, whereas elastomeric balloon pumps exhibited variable rates influenced by time and temperature, increasing the risk of medication overdose. Bolus dosing accuracy was clinically satisfactory for all pump mechanisms.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-04-19
DOI
https://doi.org/10.1038/s41598-025-98533-8
Citations
3
Primary Topic
Intravenous Infusion Technology and Safety
Type
article
Field-Weighted Citation Impact
2.02

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of infusion flow rates and bolus doses for portable infusion pump

Michiko Kinoshita, Katsuya Tanaka, Ryo Sekiguchi, Yuki Maeda
3 citations
Scientific Reports
Intravenous Infusion Technology and Safety
2.02
article

Accuracy of infusion flow rates and bolus doses for portable infusion pump

Michiko Kinoshita, Katsuya Tanaka, Ryo Sekiguchi, Yuki Maeda
article en
3 citations

Abstract

This study aimed to investigate the flow rate accuracy and bolus doses of four portable pumps with different mechanisms. We evaluated two electronic (COOPDECH Amy PCA [Amy] and CADD-Solis) and two elastomeric balloon-type pumps (COOPDECH Balloonjector and Rakurakufuser) under the following conditions: placement with an epidural catheter, operation on a shaking table, warming to 32 °C, and at room temperature (control, 25 °C). Amy maintained consistent flow rates across all conditions. The CADD-Solis also exhibited consistent flow rates, with minor yet statistically significant changes upon epidural catheter placement (- 5.0%, P = 0.005) and motion conditions (4.2%, P = 0.015). The Balloonjector and Rakurakufuser exhibited flow rate variations over time, and temperature increases significantly increased flow rates by 24.3% (P < 0.001) and 20.3% (P < 0.001), respectively. Bolus volume accuracy for the Amy, Balloonjector, and Rakurakufuser was not significantly affected under different conditions. The CADD-Solis showed a slight decrease in bolus volume with an epidural catheter (mean difference, 0.3 mL; P = 0.003). The electronic pumps maintained consistent flow rates across various conditions, whereas elastomeric balloon pumps exhibited variable rates influenced by time and temperature, increasing the risk of medication overdose. Bolus dosing accuracy was clinically satisfactory for all pump mechanisms.

Scientific ReportsVol. 15(1)
Tokushima University Hospital (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 14%
Intravenous Infusion Technology and Safety
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Accuracy of infusion flow rates and bolus doses for portable infusion pump — Michiko Kinoshita, Katsuya Tanaka, et al. · Scientific Reports (2025) | TGRS Research Map | TGRS