Performance Evaluation of the New High-Sensitivity Cardiac Troponin I Immunoassay on CL-2600i Mindray Analyzer

<b>Background</b>: International guidelines recommend the use of high-sensitivity cardiac troponin (hs-cTn) I and T methods for the detection of myocardial injury as a pre-requisite for the diagnosis of acute myocardial infarction (AMI) in patients admitted to the emergency department. Recently, Mindray (Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China) has introduced a new chemiluminescence immunoassay (CLIA) for the detection of the cTn complex. The present study aims to verify and validate the hs-cTnI Mindray assay on the new automated CL2600i analyzer compared to the routine Alinity-i series instrument by Abbott (Abbott, Chicago, IL, USA). <b>Methods</b>: This study evaluated linearity, precision through the 5 × 5 protocol, methodological comparison on plasma and serum matrices, hs-cTnI 99th percentile imprecision, and the hs-cTnI detection rate in a healthy population. <b>Results</b>: The results obtained proved that the performance of the Mindray hs-cTnI test on the CL2600i platform was closely comparable to the Abbott Alinity-i system (plasma R<sup>2</sup>: 0.974; serum R<sup>2</sup>: 0.995). The CVs were consistently low, and no significant differences were reported. Excellent analytical performance, with high sensitivity, was also observed in the healthy population (overall detection rate: 79%), as well as good linearity within the measuring range (R<sup>2</sup>: 0.994). <b>Conclusions</b>: The Mindray hs-cTnI test confirms its robustness and utility in routine practice as an advanced assay. The new technology, with more sensitive detection methods, may improve the accuracy and reliability of cardiac biomarker testing, ultimately leading to better outcomes in the management of patients with AMI and other cardiac conditions.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-04-18
DOI
https://doi.org/10.3390/diagnostics15081031
Primary Topic
Acute Myocardial Infarction Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Evaluation of the New High-Sensitivity Cardiac Troponin I Immunoassay on CL-2600i Mindray Analyzer

E. Cappa, Alfredo Giovannelli, Sergio Bernardini, Flaminia Tomassetti et al.
Diagnostics
Acute Myocardial Infarction Research
article

Performance Evaluation of the New High-Sensitivity Cardiac Troponin I Immunoassay on CL-2600i Mindray Analyzer

E. Cappa, Alfredo Giovannelli, Sergio Bernardini, Flaminia Tomassetti, Marco Alfonso Perrone, Martina Pelagalli, Massimo Pieri, Marilena Minieri, Eleonora Nicolai, Denise Fiorelli, Silvia Velocci
article en

Abstract

<b>Background</b>: International guidelines recommend the use of high-sensitivity cardiac troponin (hs-cTn) I and T methods for the detection of myocardial injury as a pre-requisite for the diagnosis of acute myocardial infarction (AMI) in patients admitted to the emergency department. Recently, Mindray (Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China) has introduced a new chemiluminescence immunoassay (CLIA) for the detection of the cTn complex. The present study aims to verify and validate the hs-cTnI Mindray assay on the new automated CL2600i analyzer compared to the routine Alinity-i series instrument by Abbott (Abbott, Chicago, IL, USA). <b>Methods</b>: This study evaluated linearity, precision through the 5 × 5 protocol, methodological comparison on plasma and serum matrices, hs-cTnI 99th percentile imprecision, and the hs-cTnI detection rate in a healthy population. <b>Results</b>: The results obtained proved that the performance of the Mindray hs-cTnI test on the CL2600i platform was closely comparable to the Abbott Alinity-i system (plasma R<sup>2</sup>: 0.974; serum R<sup>2</sup>: 0.995). The CVs were consistently low, and no significant differences were reported. Excellent analytical performance, with high sensitivity, was also observed in the healthy population (overall detection rate: 79%), as well as good linearity within the measuring range (R<sup>2</sup>: 0.994). <b>Conclusions</b>: The Mindray hs-cTnI test confirms its robustness and utility in routine practice as an advanced assay. The new technology, with more sensitive detection methods, may improve the accuracy and reliability of cardiac biomarker testing, ultimately leading to better outcomes in the management of patients with AMI and other cardiac conditions.

DiagnosticsVol. 15(8)
University of Rome Tor Vergata (IT), Saint Camillus International University of Health and Medical Sciences (IT)
Good health and well-being
Openalex Percentile: Top 86%
Acute Myocardial Infarction Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.