A geometric calibration method for a multi-segment static CT based on ordered subsets of sources and detectors

Multi-segment static computed tomography (MS-staticCT) is a generalized and efficient configuration of static CT systems, achieving high temporal resolution imaging by sequentially firing x-ray sources, instead of rotation. However, it contains numerous geometric parameters. Due to the dense arrangement of both the x-ray sources and detectors within their respective configurations, there are some coupled illumination relationships where some x-ray sources simultaneously illuminate multiple detectors. To address these calibration challenges, we propose a geometric calibration method based on ordered subsets. We categorize two types of ordered subsets of sources and detectors: source subsets and detector subsets. Each source subset includes a group of sources that illuminate the same detectors, along with the illuminated detectors. Similarly, each detector subset includes a group of detectors illuminated by the same sources, along with the sources that illuminate them. The calibration of the sources in source subsets and the detectors in detector subsets is performed alternately until convergence, ensuring that the calibrated geometry to accurately describe all the illumination relationships. These calibration steps are detailed in a workflow. During each step, the estimations for different ordered subsets are independent and parallelizable to significantly improving computational efficiency. A calibration phantom is involved in our method. During the calibration, we iteratively estimate the parameters by minimizing the average re-projection error (aRPE) of the balls in the calibration phantom. We evaluated the proposed method by simulation and actual experiments. The aRPE was reduced to 0.0087 mm and the reconstructed images were clear without obvious misalignment in simulation. Compared to estimating all parameters together, our method improved computational efficiency by a factor of 2.20. The targeted spatial resolution (2.5 lp·mm<sup>-1</sup>) of an actual MS-staticCT system was obtained. These results verified the efficiency and accuracy of the proposed method.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-04-17
DOI
https://doi.org/10.1088/2057-1976/adce0f
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A geometric calibration method for a multi-segment static CT based on ordered subsets of sources and detectors

Jing Li, Yuxiang Xing, Zhiqiang Chen, Changyu Chen
1 citations
Biomedical Physics & Engineering Express
Medical Imaging Techniques and Applications
1.36
article

A geometric calibration method for a multi-segment static CT based on ordered subsets of sources and detectors

Jing Li, Yuxiang Xing, Zhiqiang Chen, Changyu Chen
article en
1 citations

Abstract

Multi-segment static computed tomography (MS-staticCT) is a generalized and efficient configuration of static CT systems, achieving high temporal resolution imaging by sequentially firing x-ray sources, instead of rotation. However, it contains numerous geometric parameters. Due to the dense arrangement of both the x-ray sources and detectors within their respective configurations, there are some coupled illumination relationships where some x-ray sources simultaneously illuminate multiple detectors. To address these calibration challenges, we propose a geometric calibration method based on ordered subsets. We categorize two types of ordered subsets of sources and detectors: source subsets and detector subsets. Each source subset includes a group of sources that illuminate the same detectors, along with the illuminated detectors. Similarly, each detector subset includes a group of detectors illuminated by the same sources, along with the sources that illuminate them. The calibration of the sources in source subsets and the detectors in detector subsets is performed alternately until convergence, ensuring that the calibrated geometry to accurately describe all the illumination relationships. These calibration steps are detailed in a workflow. During each step, the estimations for different ordered subsets are independent and parallelizable to significantly improving computational efficiency. A calibration phantom is involved in our method. During the calibration, we iteratively estimate the parameters by minimizing the average re-projection error (aRPE) of the balls in the calibration phantom. We evaluated the proposed method by simulation and actual experiments. The aRPE was reduced to 0.0087 mm and the reconstructed images were clear without obvious misalignment in simulation. Compared to estimating all parameters together, our method improved computational efficiency by a factor of 2.20. The targeted spatial resolution (2.5 lp·mm<sup>-1</sup>) of an actual MS-staticCT system was obtained. These results verified the efficiency and accuracy of the proposed method.

Biomedical Physics & Engineering ExpressVol. 11(3)
Tsinghua University (CN)
National Natural Science Foundation of China
Affordable and clean energy
Openalex Percentile: Top 21%
Medical Imaging Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.