Clinical validation of a cross-platform digital visual acuity measurement system

PurposeThis study aimed to validate visual acuity measurements using a cross-platform system in an ophthalmological setting.MethodsVisual acuity was assessed at a distance of 5 m using two different modalities: printed and digital optotypes. The devices included Android, LG WebOS and Samsung Tizen smart TVs. Optotypes were presented in the logMAR scale. Two modalities, a single row and a block of symbols, presented SLOAN letters and Tumbling-E symbols.ResultsVisual acuity measurements of 190 participants aged 12 to 60 years demonstrated good-to-strong test-retest correlation (ICC > 0.75) and minimal bias (-0.03 to 0.02 logMAR). Limits of agreement were comparable to other studies (0.13-0.26 logMAR), with the smallest values for Tumbling-E row presentation in all devices and the highest value for Sloan chart in LG device. ANOVA revealed no statistical differences in visual acuity across devices. Sloan letters showed superior visual acuity compared to Tumbling-E (<i>p</i> < 0.001), however, this difference corresponded to only 2 letters of visual acuity.ConclusionThe digital cross-platform evaluated serves as a versatile substitute for traditional visual acuity assessments in individuals aged 12 years and older. Further research is necessary for patients with visual acuity worse than 0.5 logMAR and for conducting vision screenings in young children.

Authors

Publication Details

Journal
European Journal of Ophthalmology
Published
2025-04-17
DOI
https://doi.org/10.1177/11206721251336001
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical validation of a cross-platform digital visual acuity measurement system

J. Costa, Daniel Machado de Amorim, Paulo Roberto Lins
European Journal of Ophthalmology
Ophthalmology and Visual Impairment Studies
article

Clinical validation of a cross-platform digital visual acuity measurement system

J. Costa, Daniel Machado de Amorim, Paulo Roberto Lins
article en

Abstract

PurposeThis study aimed to validate visual acuity measurements using a cross-platform system in an ophthalmological setting.MethodsVisual acuity was assessed at a distance of 5 m using two different modalities: printed and digital optotypes. The devices included Android, LG WebOS and Samsung Tizen smart TVs. Optotypes were presented in the logMAR scale. Two modalities, a single row and a block of symbols, presented SLOAN letters and Tumbling-E symbols.ResultsVisual acuity measurements of 190 participants aged 12 to 60 years demonstrated good-to-strong test-retest correlation (ICC > 0.75) and minimal bias (-0.03 to 0.02 logMAR). Limits of agreement were comparable to other studies (0.13-0.26 logMAR), with the smallest values for Tumbling-E row presentation in all devices and the highest value for Sloan chart in LG device. ANOVA revealed no statistical differences in visual acuity across devices. Sloan letters showed superior visual acuity compared to Tumbling-E (<i>p</i> < 0.001), however, this difference corresponded to only 2 letters of visual acuity.ConclusionThe digital cross-platform evaluated serves as a versatile substitute for traditional visual acuity assessments in individuals aged 12 years and older. Further research is necessary for patients with visual acuity worse than 0.5 logMAR and for conducting vision screenings in young children.

European Journal of OphthalmologyVol. 35(5)
Openalex Percentile: Top 88%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Clinical validation of a cross-platform digital visual acuity measurement system — J. Costa, Daniel Machado de Amorim, et al. · European Journal of Ophthalmology (2025) | TGRS Research Map | TGRS