Abstract The uniformity calibration of a multilayer ionization chamber (MLIC) is crucial for ensuring the quality of depth dose distribution measurement and a comparative evaluation in the performance of calibration methods is lacking. In this study, we developed an MLIC and the mean water-equivalent length (WEL) of each layer was determined. The calibration coefficients of this MLIC were obtained through three different calibration methods and were used to correct the raw data of five energies. Measurements demonstrated high repeatability (average relative standard deviation ≤ 1.0% across all channels) for two key characteristic parameters of integrated depth dose (IDD): R 80 , defined as the depth at which the dose falls off to 80% of its maximum value, and dose distal fall-off (DDF), defined as the distance between the distal 80% and 20% of the maximum dose. The short-term stability of these parameters was within ± 0.01 mm for R 80 and ± 0.001 mm for DDF, respectively. These three calibration methods showed insignificant contributions in reducing differences between R 80 and DDF values measured by the MLIC and their reference values in water. Across all five energies examined, the average relative deviations achieved with a so-called “shifted board” method, water phantom method and penetration method are 9.78%, 16.72%, and 15.55%, respectively. The shifted board method demonstrates superior agreement with reference profiles measured in a water phantom, surpassing the other two calibration methods.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-04-01
DOI
https://doi.org/10.1088/1748-0221/20/04/p04020
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparative evaluation of calibration methods for a multilayer ionization chamber

Hongsong Bai, Mei Chen, Shuaizhe Gu, Deming Li et al.
Journal of Instrumentation
Radiation Therapy and Dosimetry
article

A comparative evaluation of calibration methods for a multilayer ionization chamber

Hongsong Bai, Mei Chen, Shuaizhe Gu, Deming Li, Xiufang Li, Zhiling Chen, Qi Liu, Yu-Hang Zou, Ye Lin
article en

Abstract

Abstract The uniformity calibration of a multilayer ionization chamber (MLIC) is crucial for ensuring the quality of depth dose distribution measurement and a comparative evaluation in the performance of calibration methods is lacking. In this study, we developed an MLIC and the mean water-equivalent length (WEL) of each layer was determined. The calibration coefficients of this MLIC were obtained through three different calibration methods and were used to correct the raw data of five energies. Measurements demonstrated high repeatability (average relative standard deviation ≤ 1.0% across all channels) for two key characteristic parameters of integrated depth dose (IDD): R 80 , defined as the depth at which the dose falls off to 80% of its maximum value, and dose distal fall-off (DDF), defined as the distance between the distal 80% and 20% of the maximum dose. The short-term stability of these parameters was within ± 0.01 mm for R 80 and ± 0.001 mm for DDF, respectively. These three calibration methods showed insignificant contributions in reducing differences between R 80 and DDF values measured by the MLIC and their reference values in water. Across all five energies examined, the average relative deviations achieved with a so-called “shifted board” method, water phantom method and penetration method are 9.78%, 16.72%, and 15.55%, respectively. The shifted board method demonstrates superior agreement with reference profiles measured in a water phantom, surpassing the other two calibration methods.

Journal of InstrumentationVol. 20(04)
Chinese Academy of Sciences (CN), Ruijin Hospital (CN), Shanghai Genon Biological Products (China) (CN), Shanghai Advanced Research Institute (CN), Shanghai Institute of Applied Physics (CN), Wuhan Union Hospital (CN), Union Hospital (CN), University of Chinese Academy of Sciences (CN)
Openalex Percentile: Top 88%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.