Optimization of signal and noise in x-ray phase and dark field imaging with a wire mesh

Phase differences imparted by tissue are significantly larger than attenuation differences. In addition, small angle scatter from tissue microstructure can provide a dark field signal that is complementary to attenuation and phase. Unfortunately, the low spatial coherence of clinical sources reduces phase and dark field contrast. Our method structures the beam with a single low-cost wire mesh that does not need precise alignment and relaxes the coherence requirement on the source. In addition, focusing polycapillary optics, which can be permanently attached to sources, are employed to allow for the use of high-power primary sources by increasing the phase signal after the focus. However, the coarseness of the mesh reduces the phase and dark field signal-to-noise ratio (SNR) compared with grating-based techniques, so optimization of the phase and dark-field SNR is an important consideration. Here, we consider the impact on the SNR of the distances between the mesh and the source and detector, and of x-ray tube voltages, to optimize the system.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-04-16
DOI
https://doi.org/10.1088/2057-1976/adcd7d
Primary Topic
Advanced X-ray Imaging Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of signal and noise in x-ray phase and dark field imaging with a wire mesh

Carolyn A. MacDonald, Arthur Redgate, Uttam Pyakurel, Jonathan C. Petruccelli
Biomedical Physics & Engineering Express
Advanced X-ray Imaging Techniques
article

Optimization of signal and noise in x-ray phase and dark field imaging with a wire mesh

Carolyn A. MacDonald, Arthur Redgate, Uttam Pyakurel, Jonathan C. Petruccelli
article en

Abstract

Phase differences imparted by tissue are significantly larger than attenuation differences. In addition, small angle scatter from tissue microstructure can provide a dark field signal that is complementary to attenuation and phase. Unfortunately, the low spatial coherence of clinical sources reduces phase and dark field contrast. Our method structures the beam with a single low-cost wire mesh that does not need precise alignment and relaxes the coherence requirement on the source. In addition, focusing polycapillary optics, which can be permanently attached to sources, are employed to allow for the use of high-power primary sources by increasing the phase signal after the focus. However, the coarseness of the mesh reduces the phase and dark field signal-to-noise ratio (SNR) compared with grating-based techniques, so optimization of the phase and dark-field SNR is an important consideration. Here, we consider the impact on the SNR of the distances between the mesh and the source and detector, and of x-ray tube voltages, to optimize the system.

Biomedical Physics & Engineering ExpressVol. 11(3)
University at Albany, State University of New York (US)
Openalex Percentile: Top 90%
Advanced X-ray Imaging Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.