CONVENTIONAL ESR V/S AUTOMATED TEST 1(ALIFAX) ESR: A COMPARATIVE STUDY AT TERTIARY CARE CENTER

Objective: The erythrocyte sedimentation rate (ESR) is a key test for detecting inflammation, typically measured using the manual Westergren method. Implementing automated systems like Test 1 (ALIFAX) offers advantages like quicker results and improved precision. This study compares the accuracy of the conventional Westergren method with the automated Test 1 (ALIFAX) system. Methods: A cross-sectional study was conducted at RNT Medical College, Udaipur, using 100 blood samples. ESR was measured by both the manual Westergren method and the automated Test 1 analyzer, which utilizes capillary photometry. The correlation between the two methods was evaluated using Pearson’s correlation coefficient. Results: The mean ESR values for females were 41.88 mm/h (automated) versus 44.31 mm/h (manual), and for males, 32.16 mm/h (automated) versus 35.18 mm/h (manual). The Pearson’s correlation coefficient was 0.94, indicating a strong positive correlation. Differences between the methods were minimal for ESR values <60 mm/h but larger for values >60 mm/h. Conclusion: The automated Test 1 analyzer showed strong agreement with the Westergren method, suggesting it is a reliable and efficient alternative for ESR measurement in clinical practice.

Authors

Institutions

Publication Details

Journal
Asian Journal of Pharmaceutical and Clinical Research
Published
2025-04-07
DOI
https://doi.org/10.22159/ajpcr.2025v18i4.53935
Primary Topic
Quality and Safety in Healthcare
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

CONVENTIONAL ESR V/S AUTOMATED TEST 1(ALIFAX) ESR: A COMPARATIVE STUDY AT TERTIARY CARE CENTER

Abhishek Sharma, Manoj Jain, G. P. Dwivedi, NILAM CHARANIYA
Asian Journal of Pharmaceutical and Clinical Research
Quality and Safety in Healthcare
article

CONVENTIONAL ESR V/S AUTOMATED TEST 1(ALIFAX) ESR: A COMPARATIVE STUDY AT TERTIARY CARE CENTER

Abhishek Sharma, Manoj Jain, G. P. Dwivedi, NILAM CHARANIYA
article en

Abstract

Objective: The erythrocyte sedimentation rate (ESR) is a key test for detecting inflammation, typically measured using the manual Westergren method. Implementing automated systems like Test 1 (ALIFAX) offers advantages like quicker results and improved precision. This study compares the accuracy of the conventional Westergren method with the automated Test 1 (ALIFAX) system. Methods: A cross-sectional study was conducted at RNT Medical College, Udaipur, using 100 blood samples. ESR was measured by both the manual Westergren method and the automated Test 1 analyzer, which utilizes capillary photometry. The correlation between the two methods was evaluated using Pearson’s correlation coefficient. Results: The mean ESR values for females were 41.88 mm/h (automated) versus 44.31 mm/h (manual), and for males, 32.16 mm/h (automated) versus 35.18 mm/h (manual). The Pearson’s correlation coefficient was 0.94, indicating a strong positive correlation. Differences between the methods were minimal for ESR values <60 mm/h but larger for values >60 mm/h. Conclusion: The automated Test 1 analyzer showed strong agreement with the Westergren method, suggesting it is a reliable and efficient alternative for ESR measurement in clinical practice.

Asian Journal of Pharmaceutical and Clinical Research
Rabindranath Tagore Medical College (IN)
Openalex Percentile: Top 85%
Quality and Safety in Healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.