Accuracy of smartwatches for measuring heart rate and oxygen saturation in cynomolgus macaques compared to clinical standards

Continuous monitoring of physiological parameters in non-human primates (NHPs) necessitates a precise, non-invasive, and convenient method. This study aimed to validate the use of smartwatches with integrated pulse oximetry and heart rate (HR) monitoring capabilities for use in NHPs. Currently, the clinical standard for non-invasive continuous monitoring of peripheral oxygen saturation (SpO<sub>2</sub>) in NHPs has been the use of a transmittance pulse oximeter (TPO) affixed to a location of highly vascularized tissue. In a clinical setting, HR is monitored through electrocardiogram (ECG) or associated with SpO<sub>2</sub> measurement from a TPO probe utilizing photoplethysmography technology. Challenges in obtaining precise readings with TPOs stem from technological limitations and probe placement restrictions. To address these limitations, simultaneous HR and SpO<sub>2</sub> measurements were obtained from 15 cynomolgus macaques (<i>Macaca fascicularis</i>) using the Apple Watch 7 (AW 7), Apple Watch 9 (AW 9), and a clinical-grade TPO probe with integrated optical HR measurement technology (iM70, ELAN). Arterial blood gas (ABG) analysis was used as a reference method for SpO<sub>2</sub>. We found that a TPO device significantly underestimated SpO<sub>2</sub> compared to the AW 7 and AW 9 when referenced against ABG values. Smartwatch-derived HR and SpO<sub>2</sub> measurements demonstrated good agreement and minimal bias compared to the gold standard method. Overall, the AW 7 and AW 9 exhibited good agreement with clinical reference standards for HR and good agreement with the gold standard for SaO<sub>2</sub> in sedated cynomolgus macaques.

Authors

Institutions

Publication Details

Journal
Laboratory Animals
Published
2025-04-15
DOI
https://doi.org/10.1177/00236772251318910
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of smartwatches for measuring heart rate and oxygen saturation in cynomolgus macaques compared to clinical standards

Andrew Winterborn, Katya Douchant, Meghan Tiplady
Laboratory Animals
Non-Invasive Vital Sign Monitoring
article

Accuracy of smartwatches for measuring heart rate and oxygen saturation in cynomolgus macaques compared to clinical standards

Andrew Winterborn, Katya Douchant, Meghan Tiplady
article en

Abstract

Continuous monitoring of physiological parameters in non-human primates (NHPs) necessitates a precise, non-invasive, and convenient method. This study aimed to validate the use of smartwatches with integrated pulse oximetry and heart rate (HR) monitoring capabilities for use in NHPs. Currently, the clinical standard for non-invasive continuous monitoring of peripheral oxygen saturation (SpO<sub>2</sub>) in NHPs has been the use of a transmittance pulse oximeter (TPO) affixed to a location of highly vascularized tissue. In a clinical setting, HR is monitored through electrocardiogram (ECG) or associated with SpO<sub>2</sub> measurement from a TPO probe utilizing photoplethysmography technology. Challenges in obtaining precise readings with TPOs stem from technological limitations and probe placement restrictions. To address these limitations, simultaneous HR and SpO<sub>2</sub> measurements were obtained from 15 cynomolgus macaques (<i>Macaca fascicularis</i>) using the Apple Watch 7 (AW 7), Apple Watch 9 (AW 9), and a clinical-grade TPO probe with integrated optical HR measurement technology (iM70, ELAN). Arterial blood gas (ABG) analysis was used as a reference method for SpO<sub>2</sub>. We found that a TPO device significantly underestimated SpO<sub>2</sub> compared to the AW 7 and AW 9 when referenced against ABG values. Smartwatch-derived HR and SpO<sub>2</sub> measurements demonstrated good agreement and minimal bias compared to the gold standard method. Overall, the AW 7 and AW 9 exhibited good agreement with clinical reference standards for HR and good agreement with the gold standard for SaO<sub>2</sub> in sedated cynomolgus macaques.

Laboratory AnimalsVol. 59(4)
Queen's University (CA)
Openalex Percentile: Top 93%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.