Moderately strong intraclass correlations between actigraphic and polysomnographic total sleep time and sleep efficiency in older adults with sleep disturbance
Compared to PSG, the GeneActiv actigraphy device provides reliable measurements for total sleep time and sleep efficiency, but agreement was weaker for wake after sleep onset, sleep onset latency, and the number of awakenings. The device showed consistent performance across multiple nights, suggesting good reproducibility. However, it systematically overestimated total sleep time and underestimates wake-related parameters, hence it may not fully replace PSG for detailed sleep assessments.