Critical Role of Voltage Application Points in “Analog” and “Digital” Electrospray Ionization Mass Spectrometry

In electrospray ionization (ESI) mass spectrometry (MS), an electric DC potential is often applied to a metal capillary used to infuse a liquid sample. However, in some cases, especially when employing nanoelectrospray ionization (nanoESI), it is convenient to use a nonconducting capillary for sample delivery and spraying. In these cases, the potentials can be applied, for example, using a metal union placed in the proximity of the capillary outlet or to an electrode located in the sample reservoir near the capillary inlet. The optimum potential values, which warrant high MS signals, are different in these two operational conditions. A higher potential needs to be applied when the electrode is placed further away from the capillary outlet. Moreover, sample conductivity has a strong influence on the optimum potential values. Lower potentials must be used with highly conductive electrolytes. Thus, DC voltage scans are required to determine the optimum potentials. Applying electric potential to the electrode located in the sample reservoir, rather than metal union, significantly decreases the appearance of oxidized analyte peaks. We also show that a single-polarity square AC waveform can be applied to the union or sample reservoir electrode, and if its frequency is sufficiently high, it has a similar effect as decreasing DC voltage, allowing for digital control of electrospray with square waves (by varying duty cycle). Interestingly, the liquid meniscus oscillation frequency is independent of the AC signal frequency if the frequency is sufficiently high. Applying the AC signal in certain conditions stabilizes the electrospray plume. These observations reveal the resemblance of the ESI sample line to an <i>RC</i> circuit.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2025-04-15
DOI
https://doi.org/10.1021/jasms.5c00082
Citations
1
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.76
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Critical Role of Voltage Application Points in “Analog” and “Digital” Electrospray Ionization Mass Spectrometry

Pawel L. Urban, Decibel P. Elpa, Ochir Ochirov, M.‐H. HUNG
1 citations
Journal of the American Society for Mass Spectrometry
Mass Spectrometry Techniques and Applications
0.76
article

Critical Role of Voltage Application Points in “Analog” and “Digital” Electrospray Ionization Mass Spectrometry

Pawel L. Urban, Decibel P. Elpa, Ochir Ochirov, M.‐H. HUNG
article en
1 citations

Abstract

In electrospray ionization (ESI) mass spectrometry (MS), an electric DC potential is often applied to a metal capillary used to infuse a liquid sample. However, in some cases, especially when employing nanoelectrospray ionization (nanoESI), it is convenient to use a nonconducting capillary for sample delivery and spraying. In these cases, the potentials can be applied, for example, using a metal union placed in the proximity of the capillary outlet or to an electrode located in the sample reservoir near the capillary inlet. The optimum potential values, which warrant high MS signals, are different in these two operational conditions. A higher potential needs to be applied when the electrode is placed further away from the capillary outlet. Moreover, sample conductivity has a strong influence on the optimum potential values. Lower potentials must be used with highly conductive electrolytes. Thus, DC voltage scans are required to determine the optimum potentials. Applying electric potential to the electrode located in the sample reservoir, rather than metal union, significantly decreases the appearance of oxidized analyte peaks. We also show that a single-polarity square AC waveform can be applied to the union or sample reservoir electrode, and if its frequency is sufficiently high, it has a similar effect as decreasing DC voltage, allowing for digital control of electrospray with square waves (by varying duty cycle). Interestingly, the liquid meniscus oscillation frequency is independent of the AC signal frequency if the frequency is sufficiently high. Applying the AC signal in certain conditions stabilizes the electrospray plume. These observations reveal the resemblance of the ESI sample line to an <i>RC</i> circuit.

Journal of the American Society for Mass SpectrometryVol. 36(5)
National Tsing Hua University (TW)
Affordable and clean energy
Openalex Percentile: Top 33%
Mass Spectrometry Techniques and Applications
0.76
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.