A novel method for measuring the attenuation length and the group velocity of transparent liquids in a variable length cavity

The transparency of liquid scintillators or water is an important parameter for many detectors in particle and astroparticle physics. In this work, the Cavity Enhanced Long Light Path Attenuation Length Screening (CELLPALS) method for the determination of the attenuation length is presented for the first time. The method is based on an experimental setup similar to a Fabry-Pérot interferometer but adding up multiple-reflected intensities of a modulated light source. CELLPALS was developed to measure the attenuation length of highly transparent liquids ( > 10 m ) with significantly lower uncertainties than with UV–Vis spectroscopy, which is a standard method for determining the attenuation length. In addition to the attenuation length, the group velocity of light in the sample can also be derived from the free spectral range of the cavity, which is not provided by any conventional method for determining the attenuation length. In this work, the CELLPALS method, its achievable precision and an experimental setup to demonstrate its feasibility are discussed. The attenuation lengths and group velocities of several transparent liquids were measured at wavelengths between 420 nm and 435 nm . In addition, the attenuation length and group velocity of linear alkylbenzene (LAB) samples after different purification stages and a purified LAB-based liquid scintillator were measured at a wavelength of 425 nm . The results confirmed the potential of CELLPALS to determine the attenuation length with an uncertainty of ∼ 2 % . The group velocity of light in the sample can be determined with an uncertainty of ∼ 0 . 03 % .

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2025-04-15
DOI
https://doi.org/10.1016/j.nima.2025.170507
Primary Topic
Scientific Measurement and Uncertainty Evaluation
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel method for measuring the attenuation length and the group velocity of transparent liquids in a variable length cavity

D. S. V. Bandara, Jessica Eck, Tobias Heinz, Tobias Lachenmaier et al.
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Scientific Measurement and Uncertainty Evaluation
article

A novel method for measuring the attenuation length and the group velocity of transparent liquids in a variable length cavity

D. S. V. Bandara, Jessica Eck, Tobias Heinz, Tobias Lachenmaier, Gina Grünauer
article en

Abstract

The transparency of liquid scintillators or water is an important parameter for many detectors in particle and astroparticle physics. In this work, the Cavity Enhanced Long Light Path Attenuation Length Screening (CELLPALS) method for the determination of the attenuation length is presented for the first time. The method is based on an experimental setup similar to a Fabry-Pérot interferometer but adding up multiple-reflected intensities of a modulated light source. CELLPALS was developed to measure the attenuation length of highly transparent liquids ( > 10 m ) with significantly lower uncertainties than with UV–Vis spectroscopy, which is a standard method for determining the attenuation length. In addition to the attenuation length, the group velocity of light in the sample can also be derived from the free spectral range of the cavity, which is not provided by any conventional method for determining the attenuation length. In this work, the CELLPALS method, its achievable precision and an experimental setup to demonstrate its feasibility are discussed. The attenuation lengths and group velocities of several transparent liquids were measured at wavelengths between 420 nm and 435 nm . In addition, the attenuation length and group velocity of linear alkylbenzene (LAB) samples after different purification stages and a purified LAB-based liquid scintillator were measured at a wavelength of 425 nm . The results confirmed the potential of CELLPALS to determine the attenuation length with an uncertainty of ∼ 2 % . The group velocity of light in the sample can be determined with an uncertainty of ∼ 0 . 03 % .

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1077
University of Tübingen (DE)
Deutsche Forschungsgemeinschaft
Openalex Percentile: Top 92%
Scientific Measurement and Uncertainty Evaluation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.