Calibration and evaluation of the MTS-N dosimeter for absorbed dose measurement in blood components

The use of ionizing radiation has significantly expanded, particularly in the healthcare sector, where the irradiation of blood components prevents transfusion-associated graft-versus-host disease (TA-GvHD). The ISO ASTM 51939:2017 standard establishes a minimum dose of 25 Gy for this procedure, requiring a precise and traceable dosimetric system. This study investigates the calibration of MTS-N thermoluminescent dosimeters at the Gamma Irradiation Laboratory (LIG) of CDTN, which currently uses radiochromic films for routine dosimetry. The dosimeters were calibrated using a RISØ TL/OSL DA-20 reader, and the results indicated that doses above 5 Gy caused reader saturation, distorting the TL emission curve. The introduction of attenuating filters maintained linearity up to 25 Gy, and the calibration coefficient was determined as 6.774×10⁻⁷ Gy/Count, confirming the system’s suitability for dosimetry in blood irradiation at LIG/CDTN.

Authors

Institutions

Publication Details

Journal
Brazilian Journal of Radiation Sciences
Published
2025-04-14
DOI
https://doi.org/10.15392/2319-0612.2024.2502
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Calibration and evaluation of the MTS-N dosimeter for absorbed dose measurement in blood components

Peterson Lima Squair, Thêssa Cristina Alonso, Leonardo Marques da Fonseca, Daniel Silva Calheiro et al.
Brazilian Journal of Radiation Sciences
Advanced Radiotherapy Techniques
article

Calibration and evaluation of the MTS-N dosimeter for absorbed dose measurement in blood components

Peterson Lima Squair, Thêssa Cristina Alonso, Leonardo Marques da Fonseca, Daniel Silva Calheiro, Amir Zacarías Mesquita
article en

Abstract

The use of ionizing radiation has significantly expanded, particularly in the healthcare sector, where the irradiation of blood components prevents transfusion-associated graft-versus-host disease (TA-GvHD). The ISO ASTM 51939:2017 standard establishes a minimum dose of 25 Gy for this procedure, requiring a precise and traceable dosimetric system. This study investigates the calibration of MTS-N thermoluminescent dosimeters at the Gamma Irradiation Laboratory (LIG) of CDTN, which currently uses radiochromic films for routine dosimetry. The dosimeters were calibrated using a RISØ TL/OSL DA-20 reader, and the results indicated that doses above 5 Gy caused reader saturation, distorting the TL emission curve. The introduction of attenuating filters maintained linearity up to 25 Gy, and the calibration coefficient was determined as 6.774×10⁻⁷ Gy/Count, confirming the system’s suitability for dosimetry in blood irradiation at LIG/CDTN.

Brazilian Journal of Radiation SciencesVol. 12(4A (Suppl.))
Universidade Federal de Minas Gerais (BR), Centro de Desenvolvimento da Tecnologia Nuclear (BR)
Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Conselho Nacional de Desenvolvimento Científico e Tecnológico, Fundação de Amparo à Pesquisa do Estado de Minas Gerais, Comissão Nacional de Energia Nuclear
Openalex Percentile: Top 91%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.