Calibration and evaluation of the MTS-N dosimeter for absorbed dose measurement in blood components
The use of ionizing radiation has significantly expanded, particularly in the healthcare sector, where the irradiation of blood components prevents transfusion-associated graft-versus-host disease (TA-GvHD). The ISO ASTM 51939:2017 standard establishes a minimum dose of 25 Gy for this procedure, requiring a precise and traceable dosimetric system. This study investigates the calibration of MTS-N thermoluminescent dosimeters at the Gamma Irradiation Laboratory (LIG) of CDTN, which currently uses radiochromic films for routine dosimetry. The dosimeters were calibrated using a RISØ TL/OSL DA-20 reader, and the results indicated that doses above 5 Gy caused reader saturation, distorting the TL emission curve. The introduction of attenuating filters maintained linearity up to 25 Gy, and the calibration coefficient was determined as 6.774×10⁻⁷ Gy/Count, confirming the system’s suitability for dosimetry in blood irradiation at LIG/CDTN.
Authors
- Peterson Lima Squair
- Thêssa Cristina Alonso (ORCID: https://orcid.org/0000-0003-2066-6785)
- Leonardo Marques da Fonseca (ORCID: https://orcid.org/0000-0001-8239-6843)
- Daniel Silva Calheiro (ORCID: https://orcid.org/0000-0001-6822-3065)
- Amir Zacarías Mesquita (ORCID: https://orcid.org/0000-0003-3411-5984)
Institutions
- Universidade Federal de Minas Gerais (BR)
- Centro de Desenvolvimento da Tecnologia Nuclear (BR)
Publication Details
- Journal
- Brazilian Journal of Radiation Sciences
- Published
- 2025-04-14
- DOI
- https://doi.org/10.15392/2319-0612.2024.2502
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
- Conselho Nacional de Desenvolvimento Científico e Tecnológico
- Fundação de Amparo à Pesquisa do Estado de Minas Gerais
- Comissão Nacional de Energia Nuclear