Intra‐Laboratory and Inter‐Laboratory Variations Analysis for <scp>HbA1c</scp> Assays in China: Using Internal Quality Control and External Quality Assessment Data
Both inter-laboratory and intra-laboratory variations of HbA1c measurement have significantly decreased over the years. However, the difference among manufacturers still exists, and ongoing efforts are required to fully comply with clinical guideline requirements.
Authors
- Yongyong Lou
- Fengfeng Kang (ORCID: https://orcid.org/0000-0002-1991-4801)
- Zhiming Shan (ORCID: https://orcid.org/0000-0002-4575-3933)
- Qian Chen (ORCID: https://orcid.org/0009-0003-8190-8482)
Institutions
- Zhejiang Provincial People's Hospital (CN)
- Hangzhou Medical College (CN)
- Zhejiang University (CN)
Publication Details
- Journal
- Journal of Clinical Laboratory Analysis
- Published
- 2025-04-01
- DOI
- https://doi.org/10.1002/jcla.70036
- Citations
- 5
- Primary Topic
- Blood Pressure and Hypertension Studies
- Type
- article
- Field-Weighted Citation Impact
- 6.94