The effect of inadequate drying of blood spots on newborn screening analyte concentrations

BackgroundA critical pre-analytical phase of newborn screening (NBS) testing involves the drying of blood applied to the blood collection devices to form the dried blood spots (DBS). Guidance states that blood applied should be air-dried for a minimum of 3 h. A recent survey highlighted that a number of DBS specimens routinely received into laboratories have a 'crinkled' appearance and that DBS specimens collected in a hospital setting are transported to the laboratory in sealed plastic bags. To date no scientific studies have evaluated aspects of blood drying on DBS NBS analyte concentrations.MethodsWe undertook experiments to recreate 'crinkled' DBS specimens in the laboratory and assess the impact on analyte concentrations. We also assessed the impact of storing collection devices following blood application in hermetically sealed plastic bags to impede the drying process. Experiments were performed using whole blood enriched with thyroid stimulating hormone, immunoreactive trypsinogen, phenylalanine, tyrosine, leucine, methionine, octanoyl-carnitine, decanoyl-carnitine, isovaleryl-carnitine and glutaryl-carnitine to pathophysiological concentrations.Results'Crinkled' DBS specimens produced significantly lower results (mean -15.5%, range -25.1 to -4.7%) for all analytes measured versus air-dried DBS specimens (<i>P</i> < .05). Analyte concentrations obtained from DBS specimens following storage in plastic bags before drying were significantly lower (mean -41.6%, range -60.0 to -27.6%) for all analytes measured (<i>P</i> < .05) versus air-dried DBS specimens.ConclusionResults from this study demonstrate that all DBS specimens with a crinkled appearance and those received in plastic specimen bags should be rejected and a repeat specimen collected to prevent erroneous screening results.

Authors

Institutions

Publication Details

Journal
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Published
2025-04-12
DOI
https://doi.org/10.1177/00045632251337608
Primary Topic
Blood groups and transfusion
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The effect of inadequate drying of blood spots on newborn screening analyte concentrations

L. L. Woodcock, Stuart J. Moat, Jude Kay, Sikha de Souza et al.
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Blood groups and transfusion
article

The effect of inadequate drying of blood spots on newborn screening analyte concentrations

L. L. Woodcock, Stuart J. Moat, Jude Kay, Sikha de Souza, M. Levi, Annabel Rodham, Margaret Birch, Nadia Worlock, Chandra Sundas
article en

Abstract

BackgroundA critical pre-analytical phase of newborn screening (NBS) testing involves the drying of blood applied to the blood collection devices to form the dried blood spots (DBS). Guidance states that blood applied should be air-dried for a minimum of 3 h. A recent survey highlighted that a number of DBS specimens routinely received into laboratories have a 'crinkled' appearance and that DBS specimens collected in a hospital setting are transported to the laboratory in sealed plastic bags. To date no scientific studies have evaluated aspects of blood drying on DBS NBS analyte concentrations.MethodsWe undertook experiments to recreate 'crinkled' DBS specimens in the laboratory and assess the impact on analyte concentrations. We also assessed the impact of storing collection devices following blood application in hermetically sealed plastic bags to impede the drying process. Experiments were performed using whole blood enriched with thyroid stimulating hormone, immunoreactive trypsinogen, phenylalanine, tyrosine, leucine, methionine, octanoyl-carnitine, decanoyl-carnitine, isovaleryl-carnitine and glutaryl-carnitine to pathophysiological concentrations.Results'Crinkled' DBS specimens produced significantly lower results (mean -15.5%, range -25.1 to -4.7%) for all analytes measured versus air-dried DBS specimens (<i>P</i> < .05). Analyte concentrations obtained from DBS specimens following storage in plastic bags before drying were significantly lower (mean -41.6%, range -60.0 to -27.6%) for all analytes measured (<i>P</i> < .05) versus air-dried DBS specimens.ConclusionResults from this study demonstrate that all DBS specimens with a crinkled appearance and those received in plastic specimen bags should be rejected and a repeat specimen collected to prevent erroneous screening results.

Annals of Clinical Biochemistry International Journal of Laboratory MedicineVol. 62(6)
Cardiff and Vale University Health Board (GB), University Hospital of Wales (GB), Public Health Wales (GB), Screen (US)
Openalex Percentile: Top 90%
Blood groups and transfusion
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.