Performance of the Binding Site cerebrospinal fluid kappa free light chains assay

BackgroundMeasurement of cerebrospinal fluid (CSF) kappa free light chains (KFLCs) for the detection of intrathecal immunoglobulin synthesis has generated interest as an alternative to CSF oligoclonal bands due to its rapid turnaround time and ability to automate the assay. A turbidimetric immunoassay - Freelite Mx, is available from the Binding Site, but published data on performance is scant. Therefore, we undertook a multicentre sample comparison and investigated reagent lot-to-lot-variation (LTLV).MethodsIntra-/inter-assay accuracy and imprecision of the Freelite Mx assay on the Binding Site Optilite analyser was assessed. Twenty paired CSF/serum samples were sent to three laboratories within Australia for the measurement of CSF/serum KFLC/albumin and concentrations compared using the Kruskal-Wallis test, Spearman's rank (rs), and Passing-Bablok analysis. Lot-to-lot-variation between three reagent lots was undertaken by analysis of 20 CSF samples.ResultsIntra- and inter-assay imprecision was ≤4.4 and ≤4.1%, respectively. There was a good correlation (rs = ≥0.98) between sites for the measured CSF KFLC concentration, and no significant difference in the median concentration measured between sites (3.31, 2.78, and 3.48 mg/L, <i>P</i> = .98). The median bias between reagent lots was <4%, the intercept of the regression between lots was between -0.02 and 0.06 mg/L, and the slope ranged from 0.96 to 1.07.ConclusionOverall, there was a good agreement in CSF KFLC concentrations among laboratories, and LTLV was deemed acceptable. Ascertaining biological variability of CSF KFLCs and the participation of laboratories in quality assurance schemes would assist with harmonisation.

Authors

Institutions

Publication Details

Journal
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Published
2025-04-12
DOI
https://doi.org/10.1177/00045632251337616
Citations
1
Primary Topic
Advanced Biosensing Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.81
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of the Binding Site cerebrospinal fluid kappa free light chains assay

Troy Wanandy, Simon A. Handley, S. Barnes, N.A. Jenkins
1 citations
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Advanced Biosensing Techniques and Applications
0.81
article

Performance of the Binding Site cerebrospinal fluid kappa free light chains assay

Troy Wanandy, Simon A. Handley, S. Barnes, N.A. Jenkins
article en
1 citations

Abstract

BackgroundMeasurement of cerebrospinal fluid (CSF) kappa free light chains (KFLCs) for the detection of intrathecal immunoglobulin synthesis has generated interest as an alternative to CSF oligoclonal bands due to its rapid turnaround time and ability to automate the assay. A turbidimetric immunoassay - Freelite Mx, is available from the Binding Site, but published data on performance is scant. Therefore, we undertook a multicentre sample comparison and investigated reagent lot-to-lot-variation (LTLV).MethodsIntra-/inter-assay accuracy and imprecision of the Freelite Mx assay on the Binding Site Optilite analyser was assessed. Twenty paired CSF/serum samples were sent to three laboratories within Australia for the measurement of CSF/serum KFLC/albumin and concentrations compared using the Kruskal-Wallis test, Spearman's rank (rs), and Passing-Bablok analysis. Lot-to-lot-variation between three reagent lots was undertaken by analysis of 20 CSF samples.ResultsIntra- and inter-assay imprecision was ≤4.4 and ≤4.1%, respectively. There was a good correlation (rs = ≥0.98) between sites for the measured CSF KFLC concentration, and no significant difference in the median concentration measured between sites (3.31, 2.78, and 3.48 mg/L, <i>P</i> = .98). The median bias between reagent lots was <4%, the intercept of the regression between lots was between -0.02 and 0.06 mg/L, and the slope ranged from 0.96 to 1.07.ConclusionOverall, there was a good agreement in CSF KFLC concentrations among laboratories, and LTLV was deemed acceptable. Ascertaining biological variability of CSF KFLCs and the participation of laboratories in quality assurance schemes would assist with harmonisation.

Annals of Clinical Biochemistry International Journal of Laboratory MedicineVol. 62(6)
University of Tasmania (AU), Alfred State College (US), The Alfred Hospital (AU), Royal Hobart Hospital (AU), Pathwest Laboratory Medicine (AU), Alfred Health (AU), Alfred University (US)
Openalex Percentile: Top 29%
Advanced Biosensing Techniques and Applications
0.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.