Evaluation of electrode-tissue contact using multifrequency impedance analysis and Cole-Cole model fitting

Atrial fibrillation (AF) is a common cardiac arrhythmia, and ablation is the primary treatment for patients with drug intolerance. The success of AF ablation depends on the adhesion of the catheter to the tissue. Existing electrical coupling index (ECI) and electrode-interface resistance (IR) methods based on impedance measurement to evaluate the adhesion between catheters and tissues do not explore the internal changes of the tissue during the compression process. This study introduces a new method to understand these internal changes using multi-frequency impedance combined with Cole-Cole model fitting, which is critical for accurate characterization of the contact between catheter and tissue. We used four-electrodes impedance measurement, using customized circuits and compression platform, applying 5-400 g (3.6-228.2 Pa) pressure to the bullfrog thighs to collect impedance data at frequencies of 500-100 kHz. The Cole-Cole model was then used for data fitting and analysis. The customized circuit accurately detects impedance up to 2 kΩ with less than 5% amplitude error, less than 15% phase error, and less than 6% error in model component values. Correlation analysis showed a significant linear relationship between extracellular fluid resistance and applied pressure (Pearson <i>R</i> ≈ 0.9, <i>p</i> < 0.05), indicating that extracellular fluid resistance increases with compression. This suggests that there is a significant linear positive correlation between the extracellular fluid resistance and the applied pressure, meaning that as the pressure increases, the extracellular fluid resistance correspondingly rises. This may provide a new perspective for studying the degree of catheter-tissue contact during atrial fibrillation ablation procedures.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2025-04-01
DOI
https://doi.org/10.1177/09544119251330742
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of electrode-tissue contact using multifrequency impedance analysis and Cole-Cole model fitting

Mengying Zhan, Hongwei Yao, Qingji Xie, Yu Zhou et al.
1 links
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Electrical and Bioimpedance Tomography
article

Evaluation of electrode-tissue contact using multifrequency impedance analysis and Cole-Cole model fitting

Mengying Zhan, Hongwei Yao, Qingji Xie, Yu Zhou, Yingxi Wang
article en

Abstract

Atrial fibrillation (AF) is a common cardiac arrhythmia, and ablation is the primary treatment for patients with drug intolerance. The success of AF ablation depends on the adhesion of the catheter to the tissue. Existing electrical coupling index (ECI) and electrode-interface resistance (IR) methods based on impedance measurement to evaluate the adhesion between catheters and tissues do not explore the internal changes of the tissue during the compression process. This study introduces a new method to understand these internal changes using multi-frequency impedance combined with Cole-Cole model fitting, which is critical for accurate characterization of the contact between catheter and tissue. We used four-electrodes impedance measurement, using customized circuits and compression platform, applying 5-400 g (3.6-228.2 Pa) pressure to the bullfrog thighs to collect impedance data at frequencies of 500-100 kHz. The Cole-Cole model was then used for data fitting and analysis. The customized circuit accurately detects impedance up to 2 kΩ with less than 5% amplitude error, less than 15% phase error, and less than 6% error in model component values. Correlation analysis showed a significant linear relationship between extracellular fluid resistance and applied pressure (Pearson <i>R</i> ≈ 0.9, <i>p</i> < 0.05), indicating that extracellular fluid resistance increases with compression. This suggests that there is a significant linear positive correlation between the extracellular fluid resistance and the applied pressure, meaning that as the pressure increases, the extracellular fluid resistance correspondingly rises. This may provide a new perspective for studying the degree of catheter-tissue contact during atrial fibrillation ablation procedures.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in MedicineVol. 239(4)
University of Shanghai for Science and Technology (CN)
Openalex Percentile: Top 94%
Electrical and Bioimpedance Tomography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.