<b>Background:</b> Endotracheal tube (ETT) cuffs prevent over-the-cuff secretions from flowing into the lower airways. However, they may not completely prevent fluid leakage around ETTs. To validate a cuff design with high sealing capacity, we compared ETTs with varying cuff materials and structural properties. <b>Methods:</b> We used 7 ETTs with different cuff materials (polyvinyl chloride [PVC], polyurethane [PU]), shapes (conical, tapered, or cylindrical), wall thickness, and cuff sizes (contact area with the trachea). Wall thickness was measured after cutting the cuff using a micrometer. The contact area was calculated based on the long diameter of the cuff when expanded within a clear acrylic tube. The sealing capacity was defined as the time taken for 10 mL of distilled water to leak past the cuff. The sealing capacities of the cuffs were compared by inserting them into the simulated trachea (silicone corrugated tube). <b>Results:</b> The wall thicknesses were 29-29.3 µm for PU and 45.6-285 µm for PVC cuffs. The contact areas with the trachea were 18.7-27.1 and 12.5-22.3 cm for PU and PVC cuffs, respectively. The mean (SD) sealing capacities were 2,381 (484), 437 (177), 56 (12), and 24 (4) s for the cylindrical PU, conical PU, tapered PVC, and conical PVC cuffs, respectively. For the 3 cylindrical PVC cuffs (A, B, and C, respectively), the mean (SD) sealing capacities were 53 (11), 15 (2), and 9 (2) s. <b>Conclusions:</b> The PU cuff had a higher sealing capacity than the PVC cuff, and the cylindrical cuff had a higher sealing capacity than the conical cuff. For the PVC cuff, thinner materials had higher sealing capacities. Furthermore, the contact area between the cuff and model trachea significantly affected the sealing capacity.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2025-04-12
DOI
https://doi.org/10.1089/respcare.12465
Citations
1
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
2.19
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluating the Sealing Capacities of Different Endotracheal Tube Cuff Designs

Yoshihiro Tange, Junichi Michikoshi, Kumiko Takagi, Tadashi Tomo et al.
1 citations
Respiratory Care
Airway Management and Intubation Techniques
2.19
article

Evaluating the Sealing Capacities of Different Endotracheal Tube Cuff Designs

Yoshihiro Tange, Junichi Michikoshi, Kumiko Takagi, Tadashi Tomo, Makoto Yamamoto, T Kudo
article en
1 citations

Abstract

<b>Background:</b> Endotracheal tube (ETT) cuffs prevent over-the-cuff secretions from flowing into the lower airways. However, they may not completely prevent fluid leakage around ETTs. To validate a cuff design with high sealing capacity, we compared ETTs with varying cuff materials and structural properties. <b>Methods:</b> We used 7 ETTs with different cuff materials (polyvinyl chloride [PVC], polyurethane [PU]), shapes (conical, tapered, or cylindrical), wall thickness, and cuff sizes (contact area with the trachea). Wall thickness was measured after cutting the cuff using a micrometer. The contact area was calculated based on the long diameter of the cuff when expanded within a clear acrylic tube. The sealing capacity was defined as the time taken for 10 mL of distilled water to leak past the cuff. The sealing capacities of the cuffs were compared by inserting them into the simulated trachea (silicone corrugated tube). <b>Results:</b> The wall thicknesses were 29-29.3 µm for PU and 45.6-285 µm for PVC cuffs. The contact areas with the trachea were 18.7-27.1 and 12.5-22.3 cm for PU and PVC cuffs, respectively. The mean (SD) sealing capacities were 2,381 (484), 437 (177), 56 (12), and 24 (4) s for the cylindrical PU, conical PU, tapered PVC, and conical PVC cuffs, respectively. For the 3 cylindrical PVC cuffs (A, B, and C, respectively), the mean (SD) sealing capacities were 53 (11), 15 (2), and 9 (2) s. <b>Conclusions:</b> The PU cuff had a higher sealing capacity than the PVC cuff, and the cylindrical cuff had a higher sealing capacity than the conical cuff. For the PVC cuff, thinner materials had higher sealing capacities. Furthermore, the contact area between the cuff and model trachea significantly affected the sealing capacity.

Respiratory CareVol. 70(8)
Oita University (JP), Oita University Hospital (JP)
Openalex Percentile: Top 16%
Airway Management and Intubation Techniques
2.19
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.