Feasibility of Noncontrast 3D MR Angiography on a Commercial Wide-Bore 0.55-T System: Comparison with 1.5-T MR Angiography

Purpose To evaluate a noncontrast electrocardiographically triggered, navigator-gated, three-dimensional, balanced steady-state free precession MR angiography (MRA) research sequence on a wide-bore 0.55-T commercial low-field-strength system, by comparing image quality and aortic dimension measurements against images obtained at 1.5 T. Materials and Methods Ten healthy volunteers (28.8 years ± 9.0 [SD]; four male) and 10 participants being evaluated for dilated thoracic aorta (53.6 years ± 10.7; six male) underwent noncontrast MRA scans at both 1.5 T and 0.55 T. Overall image quality, intrarater and interrater agreement in aortic dimensions, and diagnostic accuracy were evaluated between field strengths for both groups of individuals using Wilcoxon test, intraclass correlation coefficient, and Bland-Altman plots. Results Median image quality scores remained comparable between volunteers and participants across field strengths. Blood signal variability was greater at the ascending aorta at 1.5 T for volunteers (<i>P</i> = .01) and participants (<i>P</i> = .02). Blood-myocardium contrast was significantly higher at the ascending aorta and proximal arch at 1.5 T for volunteers and participants (<i>P</i> < .05). Excellent intra- and interrater agreement (intraclass correlation coefficient > 0.9) was demonstrated at 0.55 T and 1.5 T. Aortic dimensions (intraclass correlation coefficient > 0.9) and diagnosis of aortic dilatation did not demonstrate significant interfield differences. Conclusion Noncontrast, three-dimensional, balanced steady-state free precession MRA at 0.55 T demonstrated adequate image quality and no significant differences in quantitative measurements or diagnostic accuracy compared with 1.5 T. <b>Keywords:</b> Noncontrast MR Angiography, Low Field, 0.55 T, Obesity, Thoracic Aorta © RSNA, 2025.

Authors

Institutions

Publication Details

Journal
Radiology Cardiothoracic Imaging
Published
2025-04-01
DOI
https://doi.org/10.1148/ryct.240252
Citations
1
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of Noncontrast 3D MR Angiography on a Commercial Wide-Bore 0.55-T System: Comparison with 1.5-T MR Angiography

Nikita Nair, Katarzyna Gil, Orlando P. Simonetti, Matthew Tong et al.
1 citations
Radiology Cardiothoracic Imaging
Advanced MRI Techniques and Applications
1.36
article

Feasibility of Noncontrast 3D MR Angiography on a Commercial Wide-Bore 0.55-T System: Comparison with 1.5-T MR Angiography

Nikita Nair, Katarzyna Gil, Orlando P. Simonetti, Matthew Tong, Juliet Varghese, Yuchi Han, Ning Jin
article en
1 citations

Abstract

Purpose To evaluate a noncontrast electrocardiographically triggered, navigator-gated, three-dimensional, balanced steady-state free precession MR angiography (MRA) research sequence on a wide-bore 0.55-T commercial low-field-strength system, by comparing image quality and aortic dimension measurements against images obtained at 1.5 T. Materials and Methods Ten healthy volunteers (28.8 years ± 9.0 [SD]; four male) and 10 participants being evaluated for dilated thoracic aorta (53.6 years ± 10.7; six male) underwent noncontrast MRA scans at both 1.5 T and 0.55 T. Overall image quality, intrarater and interrater agreement in aortic dimensions, and diagnostic accuracy were evaluated between field strengths for both groups of individuals using Wilcoxon test, intraclass correlation coefficient, and Bland-Altman plots. Results Median image quality scores remained comparable between volunteers and participants across field strengths. Blood signal variability was greater at the ascending aorta at 1.5 T for volunteers (<i>P</i> = .01) and participants (<i>P</i> = .02). Blood-myocardium contrast was significantly higher at the ascending aorta and proximal arch at 1.5 T for volunteers and participants (<i>P</i> < .05). Excellent intra- and interrater agreement (intraclass correlation coefficient > 0.9) was demonstrated at 0.55 T and 1.5 T. Aortic dimensions (intraclass correlation coefficient > 0.9) and diagnosis of aortic dilatation did not demonstrate significant interfield differences. Conclusion Noncontrast, three-dimensional, balanced steady-state free precession MRA at 0.55 T demonstrated adequate image quality and no significant differences in quantitative measurements or diagnostic accuracy compared with 1.5 T. <b>Keywords:</b> Noncontrast MR Angiography, Low Field, 0.55 T, Obesity, Thoracic Aorta © RSNA, 2025.

Radiology Cardiothoracic ImagingVol. 7(2)
Lung Institute (US), Siemens Healthcare (United States) (US), The Ohio State University (US)
Siemens Medical Solutions USA, National Institutes of Health, National Heart, Lung, and Blood Institute
Openalex Percentile: Top 22%
Advanced MRI Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.