The Impact of Mis-Centering On CT Number Uniformity and Linearity in CT Simulator

This aim of this study is investigating the effect of mis-centering on computed tomography (CT) number uniformity and linearity in CT simulator. The current study used two phantoms: the GE water phantom for CT number uniformity evaluation, and the ACR Gammex 467 phantom for CT number linearity evaluation. The evaluations of CT number uniformity and linearity were conducted at y-axis with variations of -15, -10, -5, 0, 5, 10, and 15 mm, and with tube voltage variations of 80, 100, 120, and 140 kVp. The results demonstrated that mis-centering affects CT number uniformity. If the y-axis position increases, then CT number uniformity increases. However, mis-centering does not affect CT number linearity.

Authors

Institutions

Publication Details

Journal
International Journal of Scientific Research in Science and Technology
Published
2025-04-08
DOI
https://doi.org/10.32628/ijsrst251222613
Primary Topic
Radiomics and Machine Learning in Medical Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Impact of Mis-Centering On CT Number Uniformity and Linearity in CT Simulator

Mahfudz Fauzan, Eko Hidayanto, Ariij Naufal, Choirul Anam et al.
3 links
International Journal of Scientific Research in Science and Technology
Radiomics and Machine Learning in Medical Imaging
article

The Impact of Mis-Centering On CT Number Uniformity and Linearity in CT Simulator

Mahfudz Fauzan, Eko Hidayanto, Ariij Naufal, Choirul Anam, Riska Amilia, Moh. Shofi Nur Utami
article en

Abstract

This aim of this study is investigating the effect of mis-centering on computed tomography (CT) number uniformity and linearity in CT simulator. The current study used two phantoms: the GE water phantom for CT number uniformity evaluation, and the ACR Gammex 467 phantom for CT number linearity evaluation. The evaluations of CT number uniformity and linearity were conducted at y-axis with variations of -15, -10, -5, 0, 5, 10, and 15 mm, and with tube voltage variations of 80, 100, 120, and 140 kVp. The results demonstrated that mis-centering affects CT number uniformity. If the y-axis position increases, then CT number uniformity increases. However, mis-centering does not affect CT number linearity.

International Journal of Scientific Research in Science and TechnologyVol. 12(2)
Diponegoro University (ID), President University (ID)
Openalex Percentile: Top 89%
Radiomics and Machine Learning in Medical Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.