Machine Learning–Based Calibration of Commercial Continuous Glucose Monitoring Sensor in Nonserum Solutions: An In Vitro Validation Study

Abstract Continuous glucose monitoring (CGM) systems, such as the FreeStyle Libre Pro (Abbott Diabetes Care), offer noninvasive glucose measurement. However, their accuracy in cerebrospinal fluid (CSF) glucose monitoring remains unvalidated. This study evaluates the performance of the FreeStyle Libre sensor against a standard laboratory analyzer and proposes a regression-based calibration model to enhance measurement accuracy in neurotrauma ICU. A FreeStyle Libre sensor was integrated into an experimental setup using an adapter. Sensor readings were recorded with glucose concentrations ranging from 50 to 275 mg/dL. A standard laboratory analyzer was used as the reference. A linear regression model was trained to correct sensor deviations, with interpolation (SciPy's interp1d) used for refined predictions. Real-time data acquisition was facilitated via Universal asynchronous receiver / transmitter (UART)-based serial communication, and adaptive learning enabled model retraining upon accumulating 10 sensor laboratory value pairs. Initial sensor readings exhibited significant deviations from laboratory values, particularly at lower glucose concentrations (mean absolute relative difference [MARD]: 30.45%). Postcalibration, the MARD was reduced to 8.92%, demonstrating improved accuracy. Interpolation further minimized deviations, correcting values such as 40 mg/dL (20% deviation) to 49.1 mg/dL (1.8% deviation) and 72 mg/dL (42.4% deviation) to 123.5 mg/dL (1.2% deviation). Adaptive learning progressively reduced the root mean square error (RMSE) from 23.7 to 9.8 mg/dL after 30 updates. The calibration model makes the FreeStyle Libre sensor more accurate for CSF glucose measurements. This method might be promising for monitoring CSF glucose continuously of patients with external ventricular drainage, improving patient care in the neurotrauma ICU.

Authors

Institutions

Publication Details

Journal
Indian Journal of Neurotrauma
Published
2025-04-09
DOI
https://doi.org/10.1055/s-0045-1806943
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Machine Learning–Based Calibration of Commercial Continuous Glucose Monitoring Sensor in Nonserum Solutions: An In Vitro Validation Study

Deepak Agrawal, Megha Gautam, Aditya Choudhary
Indian Journal of Neurotrauma
Diabetes Management and Research
article

Machine Learning–Based Calibration of Commercial Continuous Glucose Monitoring Sensor in Nonserum Solutions: An In Vitro Validation Study

Deepak Agrawal, Megha Gautam, Aditya Choudhary
article en

Abstract

Abstract Continuous glucose monitoring (CGM) systems, such as the FreeStyle Libre Pro (Abbott Diabetes Care), offer noninvasive glucose measurement. However, their accuracy in cerebrospinal fluid (CSF) glucose monitoring remains unvalidated. This study evaluates the performance of the FreeStyle Libre sensor against a standard laboratory analyzer and proposes a regression-based calibration model to enhance measurement accuracy in neurotrauma ICU. A FreeStyle Libre sensor was integrated into an experimental setup using an adapter. Sensor readings were recorded with glucose concentrations ranging from 50 to 275 mg/dL. A standard laboratory analyzer was used as the reference. A linear regression model was trained to correct sensor deviations, with interpolation (SciPy's interp1d) used for refined predictions. Real-time data acquisition was facilitated via Universal asynchronous receiver / transmitter (UART)-based serial communication, and adaptive learning enabled model retraining upon accumulating 10 sensor laboratory value pairs. Initial sensor readings exhibited significant deviations from laboratory values, particularly at lower glucose concentrations (mean absolute relative difference [MARD]: 30.45%). Postcalibration, the MARD was reduced to 8.92%, demonstrating improved accuracy. Interpolation further minimized deviations, correcting values such as 40 mg/dL (20% deviation) to 49.1 mg/dL (1.8% deviation) and 72 mg/dL (42.4% deviation) to 123.5 mg/dL (1.2% deviation). Adaptive learning progressively reduced the root mean square error (RMSE) from 23.7 to 9.8 mg/dL after 30 updates. The calibration model makes the FreeStyle Libre sensor more accurate for CSF glucose measurements. This method might be promising for monitoring CSF glucose continuously of patients with external ventricular drainage, improving patient care in the neurotrauma ICU.

Indian Journal of NeurotraumaVol. 22(02)
All India Institute of Medical Sciences (IN)
Openalex Percentile: Top 90%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.