Experimental validation of LET in intensity-modulated proton therapy with a miniaturized pixel detector

<i>Objective.</i>Given the increased interest in incorporating linear energy transfer (LET) as an optimization parameter in intensity-modulated proton therapy (IMPT), a solution for experimental validation of simulations and patient-specific quality assurance (PSQA) in terms of proton LET is needed. Here, we present the methodology and results of LET spectra measurements for spread-out Bragg peak (SOBP) and IMPT plans using a miniaturized pixel detector Timepix3.<i>Approach.</i>We used a MiniPIX Timepix3 detector that provides single-particle tracking, type-resolving power, and spectral information while allowing measurement in quasi-continuous mode. We performed measurements for SOBP and IMPT plans in homogeneous RW3 and heterogeneous CIRS head phantoms with reduced beam current. An artificial intelligence-based model was applied for proton identification and a GPU-accelerated FRED Monte Carlo (MC) code was applied for corresponding MC simulations.<i>Main results.</i>We compared the deposited energy and LET spectra obtained in mixed radiation fields from measurements and MC simulations. The peak positions of deposited energy and LET spectra for the SOBP and IMPT plans agree within the error bars. Discrepancies exceeding the error bars are only visible in the logarithmic scale in high-energy deposition and high-LET tails of the distributions. The mean relative difference of dose-averaged LET values between measurements and MC simulations for individual energy layers is about 5.1%.<i>Significance.</i>This study presents a methodology for assessing radiation quality in proton therapy through energy deposition and LET spectra measurements in uniform and clinical IMPT fields. Findings show an agreement between experimental data and MC simulations, validating our approach. The presented results demonstrate the feasibility of a commercially available Timepix3 detector to validate LET computations in IMPT fields and perform PSQA in terms of LET. This will support the implementation of LET in treatment planning, which will ultimately increase the effectiveness of the treatment.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-04-09
DOI
https://doi.org/10.1088/1361-6560/adcaf9
Citations
4
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
6.07

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental validation of LET in intensity-modulated proton therapy with a miniaturized pixel detector

Antoni Ruciński, Cristina Oancea, Carlos Granja, Marzena Rydygier et al.
4 citations
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
6.07
article

Experimental validation of LET in intensity-modulated proton therapy with a miniaturized pixel detector

Antoni Ruciński, Cristina Oancea, Carlos Granja, Marzena Rydygier, Renata Kopeć, Paulina Stasica, Dawid Krzempek, J. Gajewski, Mateusz Winiarz
article en
4 citations

Abstract

<i>Objective.</i>Given the increased interest in incorporating linear energy transfer (LET) as an optimization parameter in intensity-modulated proton therapy (IMPT), a solution for experimental validation of simulations and patient-specific quality assurance (PSQA) in terms of proton LET is needed. Here, we present the methodology and results of LET spectra measurements for spread-out Bragg peak (SOBP) and IMPT plans using a miniaturized pixel detector Timepix3.<i>Approach.</i>We used a MiniPIX Timepix3 detector that provides single-particle tracking, type-resolving power, and spectral information while allowing measurement in quasi-continuous mode. We performed measurements for SOBP and IMPT plans in homogeneous RW3 and heterogeneous CIRS head phantoms with reduced beam current. An artificial intelligence-based model was applied for proton identification and a GPU-accelerated FRED Monte Carlo (MC) code was applied for corresponding MC simulations.<i>Main results.</i>We compared the deposited energy and LET spectra obtained in mixed radiation fields from measurements and MC simulations. The peak positions of deposited energy and LET spectra for the SOBP and IMPT plans agree within the error bars. Discrepancies exceeding the error bars are only visible in the logarithmic scale in high-energy deposition and high-LET tails of the distributions. The mean relative difference of dose-averaged LET values between measurements and MC simulations for individual energy layers is about 5.1%.<i>Significance.</i>This study presents a methodology for assessing radiation quality in proton therapy through energy deposition and LET spectra measurements in uniform and clinical IMPT fields. Findings show an agreement between experimental data and MC simulations, validating our approach. The presented results demonstrate the feasibility of a commercially available Timepix3 detector to validate LET computations in IMPT fields and perform PSQA in terms of LET. This will support the implementation of LET in treatment planning, which will ultimately increase the effectiveness of the treatment.

Physics in Medicine and BiologyVol. 70(9)
VSB - Technical University of Ostrava (CZ), Institute of Nuclear Physics, Polish Academy of Sciences (PL), Advacam (Czechia) (CZ)
Ministerstvo Školství, Mládeže a Tělovýchovy, Ministerstwo Edukacji i Nauki, Narodowa Agencja Wymiany Akademickiej, Narodowe Centrum Badań i Rozwoju
Affordable and clean energy
Openalex Percentile: Top 4%
Radiation Therapy and Dosimetry
6.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.