Counting multiple X-rays per pulse with an avalanche photodiode detector
Avalanche photodiode detectors (APDs) at X-ray synchrotrons are typically limited to recording at most one photon per synchrotron pulse. Digitizing the APD amplifier outputs enables signal processing to accurately measure a mean count rate of at least four photons per pulse based on initial synchrotron measurements. Higher rates are readily achievable. This method allows APDs to be utilized for time-resolved measurements at much higher intensities than before.
Authors
- Stephen M. Durbin (ORCID: https://orcid.org/0000-0002-3600-9854)
- Aidan M Jacobsen (ORCID: https://orcid.org/0009-0003-9676-9630)
- L. Powers (ORCID: https://orcid.org/0009-0004-7481-248X)
Institutions
- Purdue University West Lafayette (US)
Publication Details
- Journal
- Journal of Synchrotron Radiation
- Published
- 2025-04-09
- DOI
- https://doi.org/10.1107/s1600577525002462
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- U.S. Department of Energy
- Office of Science
- Argonne National Laboratory